AMI Deltacon DG
Program List and Explanations
77
5.3.4.2
Signal Outputs:
Select the operation mode of the signal outputs
when the relay is active:
5.3.4.3
Output/Control:
(relay or signal output):
5.3.4.4
Fault:
5.3.4.5
Delay:
Time which the instrument waits, after the input is deactivat
-
ed, before returning to normal operation.
Range: 0–6‘000 Sec
5.4 Miscellaneous
5.4.1
Language:
Set the desired language.
Available settings: German /English/French/Spanish/Italian
5.4.2
Set defaults:
Reset the instrument to factory default values in three
different ways:
Calibration:
Sets calibration values back to default. All other
values are kept in memory.
In parts:
Communication parameters are kept in memory. All
other values are set back to default values.
Completely:
Sets back all values including communication pa
-
rameters.
5.4.3
Load Firmware:
Firmware updates should be done by instructed ser
-
vice personnel only.
Continuous:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Sets the signal outputs to 0 or 4 mA. Errors, except
fatal errors, are not issued.
Continuous:
Controller continues normally.
Hold:
Controller continues based on the last valid value.
Off:
Controller is switched off.
No:
No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes:
Message E024 is issued and stored in the message
list. The Alarm relay closes when input is active.