A-96.250.491 / 190121
AMI Deltacon DG
Program List and Explanations
66
Control upwards or downwards
Setpoint:
User-defined process value for the selected parameter.
P-Band:
Range below (upwards control) or above (downwards con
-
trol) the set-point, within the dosing intensity is reduced from 100% to
0% to reach the setpoint without overshooting.
5.2.1.43
Control Parameters: if Parameters = Cond. 1(sc)
5.2.1.43.10
Setpoint
Range: 0 –3000
S
5.2.1.43.20
P-Band:
Range: 0 –3000
S
5.2.1.43
Control Parameters: if Parameters = Cond. 2(cc)
5.2.1.43.11
Setpoint
Range: 0 –3000
S
5.2.1.43.21
P-Band:
Range: 0 –3000
S
5.2.1.43
Control Parameters: if Parameters = Cond. 3(dc)
5.2.1.43.12
Setpoint
Range: 0 –3000
S
5.2.1.43.22
P-Band:
Range: 0 –3000
S
5.2.1.43
Control Parameters: if Parameters = Temp. 1
5.2.1.43.13
Setpoint
Range: -25 to +270 °C
5.2.1.43.23
P-Band:
Range: -25 to +270 °C
5.2.1.43
Control Parameters: if Parameters = Temp. 2
5.2.1.43.14
Setpoint
Range: -25 to +270 °C
5.2.1.43.24
P-Band:
Range: -25 to +270 °C
5.2.1.43
Control Parameters: if Parameters = Temp. 3
5.2.1.43.15
Setpoint
Range: -25 to +270 °C
5.2.1.43.25
P-Band:
Range: -25 to +270 °C