AMI Codes-II TC
Troubleshooting
A-96.250.631 / 280219
69
E010
Sample Flow low
– check Inlet pressure
– readjust sample flow
– clean instrument, see
E011
Temp. shorted
– check wiring of temperature sensor, see
– check temperature sensor
E012
Temp. disconnected
– check wiring of temperature sensor, see
– check temperature sensor
E013
Case Temp. high
– check case/environment temperature
– check program value
E014
Case Temp. low
– check case/environment temperature
– check program value
E015
Valve 1 defective
E016
DIS invalid
– This error appears after start-up and will
disappear after the first valid
measurement is finished.
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1 and
2
E018
Reagent Pump
– shut off power
– check wiring, see
E019
Photometer not con-
nected
– shut off power
– check wiring, see
Error
Description
Corrective action