92
A-96.250.631 / 280219
AMI Codes-II TC
Program List and Explanations
5.3.1.1.3.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value
fluctuates around the alarm value.
Range. 0.00–20.00 ppm
5.3.1.1.2.45
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0.00–28‘800 sec
5.3.1.2
Sample Temp.:
Only available if pH option is installed. Define the
measuring value, which should issue an alarm high respectively
low.
5.3.1.2.1
Alarm High:
If the sample temperature rises above the
programmed value E007 is issued. and low (E008)
Range: 30–70 °C
5.3.1.2.25
Alarm Low:
If the sample temperature falls below the programmed
value E008 is issued.
Range: 0–20 °C
5.3.1.3
Sample Flow:
Define at which sample flow a flow alarm should be
issued.
5.3.1.3.1
Flow Alarm:
Program whether the alarm relay should be activated if
there is a flow alarm. The flow alarm will always be indicated in the
display, pending error list, saved in the message list and the logger.
Available values: Yes or No
NOTICE:
Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.3.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 100–600 B/s
5.3.1.3.35
Alarm Low:
If the measuring values falls below the programmed
value E010 will be issued.
Range: 5–80 B/s
5.3.1.4
Alarm pH:
Only available if pH option is installed. Define the
measuring value, which should issue an alarm high respectively
low.
5.3.1.4.1
Alarm High:
If the measured value rises above the alarm high
value, the alarm relay is activated and E033 is displayed in the
message list.
Range: 0–14 pH