62
A-96.250.631 / 280219
AMI Codes-II TC
Maintenance
6.11. Cleaning the solenoid valve
Disassemble
the solenoid
valve
The solenoid valves are mounted at the bottom of the flow cell
block. The solenoid valve should be disassembled if it does not
switch anymore or if it is clogged.
1
Switch off the instrument according to instructions in
Operation for Maintenance, p. 45
2
Loosen the nut (A).
3
Remove the solenoid coil (B) from
the valve body (C).
4
Loosen the fixing screws of the
valve body with a 2.5 mm Allen key
(D).
A
B
C
D