SV 277 PRO User Manual
51
After confirmation (
OK
) the pop-up box closes, and the selection will be presented in the
line of the
TRIGGER
button.
Configuring Actions
To create new action, click the
field
and in the
ADD EVENT ACTION
pop-up box,
click the action you wish to add and to
configure:
Marker
,
Audio
,
I/O alarm
,
SMS alarm
or
E-mail alarm
.
After occurrence of the event, actions will be
performed during the time the event is active,
at its beginning or at the end depending on the
action type.
The
Marker
action registers a marker in the
logger file.
The Marker can be
Point
or
Block
type. The
Point marker means, that it will be registered
only at the beginning of the Event. The Block
marker will be registered at the beginning and
at the end of the Event.
The
Audio
action starts an audio recording in
the WAV file (Wave recording).
Max. duration
defines maximum time of
recording after the event start.
Min. break
defines the minimum time break
between two consequent records.
Pre trigger
enables earlier start of recording
with respect to the event start:
None
,
1 s
or
2 s
.
The
I/O Alarm
action starts an alarm signal at
the MULTI I/O socket output to which some
alarm device can be connected (for example,
alarm lamp).
Max. duration
defines maximum time of
alarm signal after the event start.
Min. break
defines the minimum time break
between two consequent alarms.