![Svantek SV 277 PRO User Manual Download Page 31](http://html1.mh-extra.com/html/svantek/sv-277-pro/sv-277-pro_user-manual_1409267031.webp)
SV 277 PRO User Manual
31
Note: Newly purchased stations have factory settings.
3. Set
Network
to
GPRS
(
path: <Menu> / Instrument /
Wireless Connection / Network
).
4. Check the Access Point Name (APN).
The default setting for the APN is "
internet
". It is
possible that your Internet provider uses different
APN. In this case the APN must be entered manually
(
path: <Menu> / Instrument / Wireless Transfer /
Modem Connection
).
5. Successful connection to SvanNET is indicated by the
“S” icon on the SVAN 977 display.
The
“S” icon means that connection
with SvanNET is established. Bars
and arrows inform about signal quality and data
transmission
– in and out.
6. To access SvanNET, log-in to
your account at:
https://www.svannet.com/panel-
login.php
Before logging, select your language.
Once logged in you can use the web
interface to work remotely with the
monitoring station.
Note: Ask your local SVANTEK distributer to create the SvanNET account for you
and assign your new station to your SvanNET account.
4.2. I
NTERFACE FUNCTIONALITIES OF MOBILE MODEM
The mobile modem enables the user a wide spectrum of interfacing capabilities using
Internet.
Internet access, provided by mobile operators, comes with a set of parameters that define
the visibility of the SIM card address in the network. This address can be
public
or
private
(accessible from the outside or not) and can be
static
or
dynamic
(same or different each
time the card connects to the Internet). The set of parameters define the spectrum of