![Svantek SV 277 PRO User Manual Download Page 50](http://html1.mh-extra.com/html/svantek/sv-277-pro/sv-277-pro_user-manual_1409267050.webp)
SV 277 PRO User Manual
50
If you click the
TRIGGER
button, the
TRIGGER CONDITIONS
configuration box
will pop-up.
In this box, you can add the condition type:
Threshold
.
The
Threshold
condition can be
of
Level+
or
Level-
type (
Mode
).
The condition is fulfilled during
the period in which the controlled
value of the selected result
(
Source
) will be higher/lower
than the
Threshold
level.
You can select as a
Source
different results
(
Leq
,
Lpeak
,
Lmax
,
Lmin
, etc. in the Sound
mode or
PEAK
,
MAX
,
RMS
, etc. in the
Vibration mode) for three profiles (
Pr 1
,
Pr 2
and
Pr 3
), measured by:
1s
, integration period
(for example,
1m, SR
) or time-history step (for
example,
50, TH
).
In the right-hand example,
1m, SR
means that
the Integration period for the Summary results
was set to 1 minute, while
50, TH
means that
the Step for the time-history was set to 50
milliseconds.
If you expand the
Threshold
by
clicking the
field, you
can define
Start
and
Stop
threshold
levels
and
delays
(
Min. duration
) of start and stop.