SV 277 PRO User Manual
11
The controller is fixed in the station case by a Signal
Power Combo connector and can also be removed
from the case. To do this the user should pull the
controller up and remove it from the slot.
To put it back the user should put the controller in the
slot and press on it to achieve a good fixation with the
connector.
Note: Only SVAN 977 and the
controller can be disconnected and removed
from the station case by the user. All other disassembling work should be
performed strictly by an authorized service team.
Note: The producer does not recommend removing the controller without a
sound reason. Double check that the controller has a good fixation in the
connector after reconnecting!
The monitoring station case is equipped with the air
pressure compensation valve that enables the user to
open the case easily if the internal pressure is lower
than the atmospheric one.
Any remaining moisture after closing the lid will be
absorbed by the silica gel bags (optional accessory).
The valve should be released if opening the top lid is
problematic.
Note: The valve must be closed when the station is used outdoors, otherwise
the case is not sealed against moisture.
The station case is equipped with three Souriau UTO type connectors for:
•
input signal from the microphone (
INPUT
).
•
powering station and charging the internal
battery (
DC SUPPLY
).
•
external interface for the meteo module or the
alarm lamp (
EXTERNAL INTERFACE
).