SV 104IS USER’S MANUAL______________________________________________________
37
4.9
Before and after measurement run
Before starting a measurement ensure:
1
the instrument is turned on (see chapter 4.3)
2
there is sufficient battery operating life and free memory by checking the status screen
(see chapter 4.4)
3
required configuration setup is selected (see chapter 4.6)
4
the
SV 104IS
is calibrated, because it affects the results (see chapter 4.7)
5
the windshield is put on because it protects the microphone from industrial environment such
as dust and moisture or from effects of impact (see chapter 3.3)
After stopping the measurement run, make sure:
1
the calibration is still maintained (see chapter 4.7)
2
the data is properly downloaded to PC for further analysis (see chapter 5)
3
the instrument is turned off (see chapter 4.3)
4.10
Starting and stopping measurement run
START:
To start the measurements the user has to press the
<SCROLL>
and
<PROFILE>
keys at
the same time. The results of the measurement are displayed in the last used result’s display view mode.
As an example ONE RESULT view mode is displayed. ONE RESULT view mode is always available for
most functions of the instrument. The results of the measurements can also be presented in other view
modes, which can be enabled or disabled
and adjusted to the user’s needs.
Figure 4-17
One profile mode screen view
STOP:
The same combination of keys:
<SCROLL>
and
<PROFILE>
allows the user to stop
the measurement run. All run results are always saved automatically, there is no need to save them
manually.
4.11
Auto-run mode information
Note, that when auto-run mode (timer and/or pause) is configured, there is information on the screen
available for the user. There is no need to turn on the instrument manually. All the timer procedures can be
easily pre-programmed with the use of the
SUPERVISOR
software.
Figure 4-18
Auto-run mode
– timer and programmable pause screen