4.4
Battery check
Observe the battery icon in the instrument’s icon status bar or press the
<ENTER>
key until
the Instrument Status
VIEW
mode is presented and check the battery state. If it is too low, charge
the batteries (chapter 4.1).
Figure 4-3 Instrument Status - Battery state
Battery status screen is moved down and up with
<SCROLL>
,
<PROFILE>
keys.
Press
<ENTER>
to change to the next
VIEW
mode.
Notice:
Battery state of charge calculation is based on internal charge counter and should
be considered only as an overall, not very precise estimation. Therefore the time left could be noticeably
different. Although the newest technology cells are used the slight degradation over time is inevitable
thus aging requires occasionally factory based (or authorised service centre) replacement of the
rechargeable battery cells.,
Notice:
Battery power indicator - To improve accuracy of remaining battery life indicator, run
the dosimeter until it is fully discharged; then proceed with a full charge via the docking station.
The procedure is recommended before first use. Repeat this procedure every few months of use