SV 104IS USER’S MANUAL______________________________________________________
39
4.13
Mounting and positioning the SV 104IS
Unless specified by local legislation, personal noise dosimeters should always be mounted on
the shoulder, circa 10 cm from the most exposed ear, with the microphone approximately a few cm above
the shoulder. SV 104IS shape and microphone height ensures proper instrument position. see Figure 4-20
below.
Notice:
The dosimeter should be mounted on conductive clothes and/or with direct touch to
human body. This assures a path to remove the accumulated electrostatic charge that could built-up on
dosimeter.
Figure 4-20
SV 104IS Positioning