18
TRIO MP-235 THREE-AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.23 (20190130) (FW V2.2 & 2.3)
3.2
Display
3.2.1
Initial Startup
Figure 3-1. LCD Display showing startup screen.
When starting the TRIO MP-235 system for the first time or if the HOME position has not
yet been defined (saved), the values of all four axes will be 1,000 micrometers (microns).
Figure 3-2. Startup screen
Figure 3-3. Factory default startup (Home) position
3.3
Control Operations
3.3.1
Maximum Positive Position Values:
Move the dial of an axis clockwise until its position value stops incrementing. The following
table lists the maximum position value (in microns) for each axis.
Table 3-1. Maximum positive position value of each axis
Axis Maximum
Position
Value (in microns)
X 25,000
Y 25,000
D 50,000
Figure 3-4. Maximum positive values
3.3.2
Setting Position for HOME or WORK
To set position, hold down HOME or WORK button for 3 seconds until beep sounds.
SUTTER
INSTRUMENT CORP.
(Text in
Green)
TRIO / MP-235
REV 2.2
(Text in
Green)
(Text in
Green)
X=25000 Y=25000
D=50000 >=
(Text in
Green)
X= 1000 Y= 1000
D= 1000 >=