15
TRIO MP-235 THREE-AXIS MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.23 (20190130) (FW V2.2 & 2.3)
2.5.2.3
Switch 6: Sensor Test (Firmware < v2.2)
Table 2-3. Configuring the Sensor Test.
Sensor Test
Switch #
Enabled **
Disabled
6
OFF (Up)
ON (Down)*
* Factory default (do not change unless requested to by Sutter Instrument Technical
Support).
** CAUTION: To avoid damage to the micromanipulator or stage, DIP Switch 6 (Sensor
Test) must always be set to OFF (DOWN).
2.5.2.4
Switch 6: Calibration Homing on Power On (Firmware v2.2+)
Configures whether calibration homing occurs or not on power on (FW v2.2+).
Table 2-4. Calibration Homing on Power On (Firmware v2.2+).
Calibration Homing on Power On
Switch #
None (No calibration)
Calibration occurs
6
OFF (Up)
ON (Down)*
* Factory default (recommended normal operation setting).
2.5.2.5
Switches 7 through 10
These four switches are reserved for all controllers programmed with a firmware version
below v2.2. For firmware v2.2 and above, the switches are defined in the following
paragraphs.
2.5.2.6
Switch 7: Speed of Manual Movement (Firmware v2.2+)
This switch determines the speed at which movement takes place when manipulated by the
ROE rotational knobs for any axis. The OFF (up) position configures the speed to “normal”
(factory default). Set the switch to ON (down) for slower speed in situations where the
micromanipulator or stage is burdened with heavier loads. This speed setting affects only
manual movement using the rotation knobs of the ROE; moves initiated by external
command (see Chapter 4) are not affected.
Table 2-5. Speed of manual movement (Firmware v2.2+).
Speed of Manual Movement
Switch #
Normal Slower**
7 OFF
(Up)*
ON (Down)
* Factory default (recommended normal operation setting).
** Use the ON (down) setting for slower speed when the micromanipulator or stage is
burdened with heavier loads.