MP-285 SYSTEM ROE BASIC OPERATIONS – REV. 4.04 (20210122)
13
To change resolution from one value to the other, simply press the <COARSE/FINE> button
on the top of the ROE (coarse is up, fine is down). The movement screen also displays the
resolution in the lower right corner “0.2 (or 0.04)
µ
m Res”. Given a resolution in microns, we
can now re-specify the amount of movement per ROE turn in microns for Setups 1 through
4:
Table 3-2. Microns per ROE turn settings
Setup
Number
Microns per ROE turn at
0.2
µ
m
0.04
µ
m
1
716.8
143.36
2
1024
204.8
3
512
102.4
4
102.4
20.48
Thus, you have a relatively wide range of precision of movement. In practice, many users end
up only using a very non-precise, fast moving setup (#1 or 2, coarse) for quickly positioning
the pipette and then a precise setup (#4, fine) for micromanipulation under the microscope.
You may find that none of these SETUPS delivers the type of responsiveness that you desire
from the MP-285. The
MP-285 Reference Manual
provides complete instructions for defining
and saving your own Setup.
Summary of Contents for MP-285
Page 3: ......
Page 4: ......
Page 8: ...MP 285 SYSTEM ROE BASIC OPERATIONS REV 4 04 20210122 vi This page intentionally left blank...
Page 12: ...MP 285 SYSTEM ROE BASIC OPERATIONS REV 4 04 20210122 2 This page intentionally left blank...
Page 20: ...MP 285 SYSTEM ROE BASIC OPERATIONS REV 4 04 20210122 10 This page intentionally left blank...
Page 24: ...MP 285 SYSTEM ROE BASIC OPERATIONS REV 4 04 20210122 14 This page intentionally left blank...
Page 34: ...MP 285 SYSTEM ROE BASIC OPERATIONS REV 4 04 20210122 24 This page intentionally left blank...
Page 36: ...MP 285 SYSTEM ROE BASIC OPERATIONS REV 4 04 20210122 26 This page intentionally left blank...
Page 38: ...MP 285 SYSTEM ROE BASIC OPERATIONS REV 4 04 20210122 28 This page intentionally blank...
Page 40: ...MP 285 SYSTEM ROE BASIC OPERATIONS REV 4 04 20210122 30 NOTES...