MP-225A MOTORIZED MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.04 (20211111)
24
D.2.
Configuration
Rear-Panel DIP Switches
Table D-1.
MP-225A configuration switches (rear).
Switch
#
Definition
State
Setting Position
(rear
view)
1
X-Axis Knob Rotation for
Forward (+) Movement
Counter
ON
UP
Clockwise* OFF* DOWN*
2
Y-Axis Knob Rotation for
Forward (+) Movement
Counter
ON
UP
Clockwise* OFF* DOWN*
3
Z-Axis Knob Rotation for
Forward (+) Movement
(Normal Mode)
Counter
ON
UP
Clockwise* OFF* DOWN*
4
Z-Axis Knob Rotation for
Forward (+) Movement
(Diagonal Mode)
Counter
ON
UP
Clockwise* OFF* DOWN*
5
Y Axis Lock Out for
Homing
Disabled*
ON*
UP*
Enabled
OFF
DOWN
6
Calibration Homing on
Power On **
Enabled *
ON*
UP*
Disabled
OFF
DOWN
NOTES:
Configuration
DIP Switches
Power Switch
Ground/-
Earth Post
Power Input
Socket
Manipulator cable
connector