12
MP-225A MOTORIZED MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.04 (20211111)
Setting a new angle for Diagonal Movements: Press and hold down SET and press and
release NORM/DIAG *(angle). The display will say “Select the angle in use (0-90)” and show
the currently saved angle (30 degrees initially). You will have 10 seconds to select a new angle
by rotating the Z axis knob. The screen will be green during this procedure.
3.2
Detailed Display Instructions
3.2.1
Initial Startup
Figure 3-2. LCD Display showing startup screen.
When starting the MP-225A system for the first time or if the HOME position has not yet
been defined (saved), the values of all four axes will be 1,000 micrometers (microns).
Figure 3-3. Startup screen
Figure 3-4. Factory default startup (Home) position with calibration enabled.
Normal screen color: Green. Move in progress (HOME, WORK, etc.), screen goes Blue, then
back to green when done.
3.3
Operation of MP-255A Integrated ROE/Controller
3.3.1
Maximum Positive Position Value
Move the dial of an axis clockwise until position value stops incrementing. The following
table lists the maximum position value (in microns) for each axis.
Table 3-1. Maximum positive position value of each axis
Axis
Maximum Position
Value (in microns)
X
25,000
Y
25,000
Z
25,000
SUTTER