LAMBDA 721 OPERATION MANUAL – REV. 1.10C (20210517)
iii
DISCLAIMER
The Lambda 721 is an ultra-high-speed wavelength switching multi-LED-based beam
combined illumination or light source device. The purpose of the system is to be an
illuminator for microscopes. No other use is recommended.
This instrument is designed for use in a laboratory environment. It is not intended, nor
should it be used in human experimentation or applied to humans in any way. This is not a
medical device.
Unless otherwise indicated in this manual or by Sutter Instrument Technical Support for
reconfiguration, do not open or attempt to repair the instrument.
Do not allow unauthorized and/or untrained operative to use this device.
Any misuse will be the sole responsibility of the user/owner and Sutter Instrument Company
assumes no implied or inferred liability for direct or consequential damages from this
instrument if it is operated or used in any way other than for which it is designed.
SAFETY WARNINGS AND PRECAUTIONS
Electrical
Operate the Lambda 721 using 110 – 240 VAC., 50-60 Hz line voltage. This instrument is
designed for use in a laboratory environment that has low electrical noise and mechanical
vibration. Surge suppression is recommended at all times.
Fuse Replacement: Replace only with the same type and rating:
5 x 20 mm glass tube, Time Lag (IEC 60127-2, Sheet 6)
T3.15A 250V (Time Lag, 3.15 Amps, 250 Volts)
(Littelfuse 02193.15MXAP (219XA Series, RoHS compliant))
A spare fuse is located in the power input module.
Avoiding Electrical Shock and Fire-related Injury
Always use the grounded power supply cord set provided to connect the system’s
power adapter to a grounded/earthed outlet (3-prong). This is required to protect you
from injury in the event that an electrical hazard occurs.
Do not disassemble the system. Refer servicing to qualified personnel.
To prevent fire or shock hazard do not expose the unit to rain or moisture.
Electromagnetic Interference
To comply with FDA and CE/EU electromagnetic immunity and interference standards; and
to reduce the electromagnetic coupling between this and other equipment in your lab always
use the type and length of interconnect cables provided with the unit for the interconnection
of its components (see the Technical Specifications appendix for more details).
Operational
Failure to comply with any of the following precautions may damage this device.
Summary of Contents for Lambda 721
Page 3: ......
Page 4: ...LAMBDA 721 OPERATION MANUAL REV 1 10C 20210517...
Page 12: ...LAMBDA 721 OPERATION MANUAL REV 1 10C 20210517 x This page intentionally blank...
Page 40: ...LAMBDA 721 OPERATION MANUAL REV 1 10C 20210517 28 This page intentionally blank...
Page 44: ...LAMBDA 721 OPERATION MANUAL REV 1 10C 20210517 32 This page intentionally blank...
Page 60: ...LAMBDA 721 OPERATION MANUAL REV 1 10C 20210517 48 NOTES...