Sutron Corporation Satlink Operations & Maintenance Manual, Rev 8.04.2 11/3/2016 pg. 161
To change the current reading of a sensor, type M1 = 12.5. This will change the
Measurement
Offset
such that the said measurement reads the user entered value. For example, if my water
level sensor, once set up, read 3.50, and I knew the water level was at 1.50, I would type M1 =
1.50. The next it was measured, M1 would read 1.50 (assuming the level does not change). The
Measurement Offset
would have changed from 0 to -2.00.
17.6.
Recording
To start Satlink, click the Start button in LinkComm, or type RECORDING = ON.
17.7.
Downloading the Log
Satlink will save the measured data in its flash memory each time a measurement is made. The
command LOG command will start a transfer of the log to the connected device.
Use NH to exclude the header; Use NE to omit events.
Help Log gives details on downloading the log.
>HELP LOG
DOWNLOAD or LOG will transfer the log from the unit
LOG ALL gets whole log, just LOG gets since last download,
LOG X gets X last days ("LOG 3" gets last 3 days worth of data)
LOG timeStart gets data since provided date
LOG timeStart timeEnd gets data between provided dates
time can be YYYY/MM/DD HH:MM:SS or YYYY/MM/DD or HH:MM:SS
eg. "LOG 12:00:00 13:00:00"
eg. "LOG 2006/01/20 12:00:00 2006/01/21 12:00:00"
append M1 to include only data from measurement one
eg. "LOG 3 M1 M2
append NH to omit log header, NE to omit events
append NY to download without YModem.