ATEC
®
Breast Biopsy and Excision System:
Sapphire
™
Unit
Operator’s Manual
Field Replaceable Items
(cont.)
Item
Instructions
Power Cord
Removal (with the console powered off)
Unplug console power cord from
wall outlet.
Completely unwrap and remove power cord
from cord management plate.
With a #0 Philips screwdriver loosen fastener
located on bottom of
retaining bracket.
Disconnect power cord connector from
console receptacle.
Installation (with the console powered off)
Insert power cord connector into console
receptacle.
With a #0 Philips screwdriver tighten
fastener located on bottom of
retaining bracket.
Wrap the power cord clockwise (CW) around
cord management plate.
Test/Verifi cation of Installation
Plug power cord into wall outlet.
Turn console power on.
Verify console powers on, beeps, and
defaults to the biopsy mode.
1.
2.
3.
4.
1.
2.
3.
1.
2.
3.
Cord
Management
Plate
Power Cord
25
Removal (with the console powered off)
1) Unplug console power cord from wall
outlet.
2) Completely unwrap and remove power
cord from cord management plate.
3) With #1 Philips screwdriver, loosen fastener
located on bottom of retaining bracket.
4) Disconnect power cord connector from
power receptacle.
Installation (with the console powered off)
1) Firmly and completely insert power
cord connector into power receptacle.
2) With #1 Philips screwdriver, tighten
fastener located on bottom of retaining
bracket.
Verifi cation of Installation
1) Plug power cord into wall outlet.
2) Turn console power on.
3) Verify console powers on, beeps
and defaults to the biopsy mode.
4) Unplug power cord from wall outlet
and wrap clockwise around cord
management plate.
Power Cord
Connector
Retaining
Bracket
Fastener