6.3.3 Load Calibration (LoAd)
After doing Data Calibration, if there are conditions for Load Caliration, do Load Calibration with
loading standard weight on the weigher to correct Span Coefficient further for higher weighing accuracy.
The loading weight should be bigger than 50% of Scale Capacity value.
【
■
】
: Exit
【
】
: Save
M
Main Display Interface
【
■
】
+
【
►
】
【
▲
】
【
】
+
【
►
】
【
▲
】
【
】
M
【
】
Input
‘Calibrating Weight’:
1~99999 Weight Unit.
[105] Span Coefficient value.
If AD Value
≤
Zero Value, display
‘
Err
’
, and it
’
s not allowed to save Span
Coefficient.