6.2 F1.SET Parameter Setting
6.2.1 Weighing Parameters (SCAL)
No.
Sign
Range
Default
Description
Set
100
Ut
0~3
1
Weight Unit
0: None
1: kg
2: t
3: g
101
dt
0~4
ooo.oo
Decimal Point
0: ooooo
1: oooo.o
2: ooo.oo
3: oo.ooo
4: o.oooo
102
SL
1~99999
10000
Scale Capacity
Max. allowed loading weight of the load receptor.
Scale Capacity
≤
(Loadcell Capacity × Loadcell
Quantity)
–
Self-weight of Load Receptor.
103
dV
1~500
1
Scale Division
1, 2, 5, 10, 20, 50, 100, 200, 500
104
ZE
-99999~
+99999
0
[*]
Zero Value
Only for query.
105
SP
>0
1.0000
[*]
Span Coefficient
Max. display value: 9.9999.
Only for query.
106
Sr
0~5
0
Stablity Judging Range
[Division]
Set value = 0: No
‘Stablity Judging’.
Set value > 0: Weight Variance per [107]
‘
Stablity
Judging Time
’
being in [106]
‘
Stablity Judging
Range
’
means
‘
Weight is stable
’
.
107
St
0.1~9.9
0.5
Stablity Judging Time
[s]
[*]:
‘
RAM Reset
’
operation has no effect on the parameter.