7.5 ORP Calibration
Under ORP measurement mode, enter calibration setup menu. Select Calibration icon,
and adjust mV value. The adjustable range is from -300mV to 300mV.
Put the ORP electrode into
ORP standard solution. Press
or to adjust the
main display value until it is
equal to the desirable mV
value.
Press to confirm it.
46