26

Press

   

for 

Press

   

for confirmation 

Press 

 

or    to 

decide the current 
output mode 
0-20 or 4-20 mA 

 

Select 0 or 4mA for testing

 

Press      or          to make test 
on 0, 4 or 20mAcurrent output 

Enter Date and Time 

Press

   

for confirmation 

Press

   

for confirmation 

Press  

 

or   to 

set the 

relative upper limit of 

Note: The setting in 7.

9

 is for PC-3100 only.

 

7.9 Analog output 2 (temperature) (PC-3100 only)   

The user can adjust the relative relationship between the TEMP measurement 

range and the output current according to actual situation, in order to improve the 
recognition of current output.   

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

 
 
 
 
 

 
 

 
 
 
 
 
 
 
 

 

 

 

Press

   

 

for confirmation 

Press

   

 

for confirmation 

Summary of Contents for PC-3100

Page 1: ...Operation Manual PC 3100 3100RS Microprocessor pH ORP Controller...

Page 2: ...l function 11 4 3 Description of terminal function 12 4 4 Installation of transmitter PH 300T 12 4 5 Connection of controller PC 3100 and transmitter PH 300T 13 4 6 Typical wirings 14 4 7 Illustration...

Page 3: ...Calibration Block diagram of Calibration 29 8 1 Security code of calibration 30 8 2 Entry of calibration mode 31 8 3 Asymmetry Buffer calibration 32 8 4 Standard Buffer calibration 33 8 5 ORP calibra...

Page 4: ...ash time and stop duration In the pH interface please set analog output 1 with or In the tP interface please set analog output 2 with or PC 3100 only In the Real time clock interface please set year d...

Page 5: ...ocedures as Step 1 Note 1 Push when showing CA2 you may choose to make single point calibration only Note 2 Single point calibration means that you only modify zero point OS value and do not modify th...

Page 6: ...lectrode completely and put it into the second buffer solution Then push bottom to start the second point calibration 4 The display will show the mV value while being calibrated After the value become...

Page 7: ...calibration steps of 1 to 7 Or push to go back to measurement mode Note 1 The dual point and three point calibration steps of pre set buffer solutions NIST are the same procedure as the calibration o...

Page 8: ...on function Analog output 1 Isolated DC 0 4 20mA corresponding to pH ORP max load 600 Analog output 2 Isolated DC 0 4 20mA corresponding to Temp max load 600 Serial Interface RS 485 Contact 240 VAC 0...

Page 9: ...stead zAvoid electrical surge when using power Especially when using three phase power use ground wire correctly 3 Assembly and installation 3 1 Controller installation This controller can be installe...

Page 10: ...controller without any joint except specific junction box Connect Set up diagram of coaxial cable See the correct set up method on the left Note The black conductive rubber covering on the central sp...

Page 11: ...and then lock C Cable fixing point MG16A tightly Leave H Sensing electrode cable for about 12 14cm and split it carefully 6 Fix the terminal of electrode central spindle on terminal block 1 of round...

Page 12: ...nds of link distributing system 1 Two wire distributing system IN terminals Terminal No OUT terminals Terminals on controller Electrode central spindle 1 Central spindle extending wire for electrode G...

Page 13: ...ls on controller Electrode central spindle 1 Central spindle extending wire for electrode GLASS Ground Rods 2 GND GND Electrode network cable 3 Green wire extending wire for electrode REF Temperature...

Page 14: ...of pH controller PC 3100 4 1 Illustration of rear panel 4 2 Illustration of terminal function D B G D A 4 20mA GLASS REF GND TP PH ORP T POWER AC INPUT POWER TP 12V HI LO WASH 4 20mA RS485 1 2 PC3100R...

Page 15: ...tput terminal for external recorder or PLC control only applicable for PC 3100 or RS 485 output D B only applicable for PC 3100RS 4 20mA terminal G Temperature current output terminal for external rec...

Page 16: ...PH 300 terminal respectively connect to DC12V and of the controller E The S and S on transmitter PH 300 terminal respectively connect to GLASS and REF of the controller F The I and I on transmitter PH...

Page 17: ...Central Network Black Two wire distribution GND REF GLAS pH ORP Central wire Network cable Short circuit slice Surge absorber Surge absorber Surge absorber Surge absorber Transmitter Washing device D...

Page 18: ...ion mode pressing this key allows you exit Calibration mode and back to Measurement mode In the parameter set up mode and Calibration mode it is an optional function key and upward key In the paramete...

Page 19: ...light HI Controlling operation indicator lamp when the high setting value is started up the REL1 indicator will light LO Controlling operation indicator lamp when the low setting value is started up t...

Page 20: ...H db1 0 10 pH Low point alarm AUTO SP2 04 00 pH db2 0 10 pH Wash time ON 0000 s OFF 000 0 H db 0000 s pH ORP current output 4 20 mA 02 00 12 00pH TP current output 4 20 mA 000 0 100 0 for PC 3100 only...

Page 21: ...reme nt setting NTC ATC OFF ATC Modify temperatur e 5 Input tempera ture PTC ATC Input REL1 hysteresis value Input REL1 setting value Relay1 Test Start up REL1 or not Yes No Relay1 Set high point Inpu...

Page 22: ...ameter Press or Press for confirmation Press for conformation Press for confirmation Press When the first 0 in the fields 0000 will shine press upward key or rightward key to input the code and press...

Page 23: ...20 7 3 Measurement parameters Enter the measurement parameter set up Enter temperature parameter setting Press or Select pH or ORP Press for Press for...

Page 24: ...y standard thermometer Press Enter High point alarm set up mode Press for Use standard thermometer to test the actual temperature of the solution and press upward or rightward press for If necessary c...

Page 25: ...the range for DB is 00 2 00pH 0 200mv Press for Press for Enter Low point alarm set up mode Press or to decide to start up REL1 or not Press or to make REL1 terminal or lamp test Choose on HI will lig...

Page 26: ...ile the range for DB is 00 2 00pH 0 200mv Press for Press for Press or to set the required alarm Press for Press for Press or to set an appropriate Press for Press for Enter automatic wash time Press...

Page 27: ...automatically stop this function Press for Press or to set the turnoff time of the washing Press for Press or to make Enter Analog output 1 pH ORP Choose on WASH lamp will light for a Press or to set...

Page 28: ...ove the recognition of current output Press or to make stimulation test for a current Press for confirmation Press or to set the relative upper limit of measurement value of 0mA Press for confirmation...

Page 29: ...confirmation Press for confirmation Press or to set the relative upper limit of Note The setting in 7 9 is for PC 3100 only 7 9 Analog output 2 temperature PC 3100 only The user can adjust the relati...

Page 30: ...for Use to set month and day The key is Press for Use to set the time The key is for digital Press for 7 11 RS 485 set up PC 3100RS only The user can set as necessry the ID and transmission speed of...

Page 31: ...Press for confirmation Use or to select 2 1 0 1 2 five backlit brightness levels Press for Back to Measurement Mode Use or to select 2 1 0 1 2 five levels of sensitivity of Press for confirmation 7 12...

Page 32: ...Cn1 rEdy Automatic Cn2 rEdy Automatic Identify the standard electrical potential Identify the standard electrical potential CA1 rEdy Manual Set pH value of standard solution Show the calibrated elect...

Page 33: ...alibration and then use or to enter the code set page Press for Press or Press for Press for Press for When the first 0 in the fields 0000 will shine press upward key or rightward key to input the cod...

Page 34: ...ue Pressing allow entering the calibration menu 8 2 4 Use or to select CA1 Ct1 or Cn1 and press for confirmation The preset is Ct1 Buffer calibration Note 1 Ct1 Refer to TECH Buffer pH2 00 pH4 01 pH7...

Page 35: ...After showing a similar number of pH value to the buffer solution push or to set the digit until it is equal to the buffer s standard Then push to ensure it 2 When the display shows CA2 push and the...

Page 36: ...r Error messages and solutions 4 The display will automatically show the oS1 zero point mV Value Push bottom to see SLP1 slope value Push again to back to measurement mode 8 4 Calibration of Pre set b...

Page 37: ...r push to back to the measurement mode Note The dual point and three point calibration steps of NTST Buffer are the same procedure as the calibration of TECH Buffer except the buffer s standard 8 4 2T...

Page 38: ...l it will show CAL Err See chapter 10 for Error messages and solutions 7 The display will show OS1 value zero point automatically Push to show the SLP1 value of calibration Push again to show OS2 valu...

Page 39: ...n Push or to set the digit until it is equal to the buffer s standard Then push to ensure it and push to back to the measurement mode 9 Instruction set for RS 485 PC 3100RS only RS 485 should apply MA...

Page 40: ...cation Hyper Terminal and the display will show a window of Hyper Terminal 2 Set icon and online name for Hyper Terminal before entering the ONLINE window 3 Select communication terminal for example s...

Page 41: ...rameter input Readout in the measurement mode and the instruction examples in set up mode 0011 refers to ID01 machine Items Instruction function Syntactic instruction Output format Data feedback insta...

Page 42: ...TS 0011TTSS 00 4 20MA 5 Read current output Relative PH ORP setting value 01RA 0011RRAA 00 2 00pH 12 00pH 6 Read current output Relative temperature setting value 01TA 0011TTAA 00 M 0 0C M 100 0C 7 Re...

Page 43: ...xxxx yyyy Refer Note1 9 Set current temperature value 01TA xxxx yyyy Refer Note2 10 WASH time setting 01WT xxxx yyyy xxxx 0 9999 sec ON yyyy 0 9999 H Off 11 High point control setting 01HM x x AUTO o...

Page 44: ...range pH mode 0 00 2 00 pH for example 0 20 pH is set as 20 ORP mode 0 200 mV for example 50 mV is set as 50 4 xxxx stands for high point alarm setting value yyy stands for high point hysteresis sett...

Page 45: ...nother calibration The readout is unstable Please check whether there is bubble or air in the glass end of the electrode maintain the electrode or change a new electrode and make another calibration T...

Page 46: ...ck Soak the electrode in Thiourea HCL solution until the electrode film turns white such as METTLER TOLEDO 9891 electrode cleaning solution Pollution by grease or organic substance Use acetone or etha...

Reviews: