38
「
Respond to the input characters (E)
」
,
「
Add LF at the end of each input line
(A)
」
and
「
newline if exceeding the width of terminal (W)
」
, and then leave
“ASCII setting” window by pushing
【
confirm
】
.
9. Push
【
confirm
】
to leave “XXX content” window.
10. Push
【
call
】
to link the communication.
9.2 RS-485 Instruction set of measurement mode:
Instruction format: ABBCC[(XXXX)]
A : Instruction leader character
BB : RS-485 ID
CC : Instruction
[] : Omission if there is no parameter
(XXXX): Parameter input
Readout in the measurement mode and the instruction examples in set-up mode <0011
refers to ID01 machine>
Items Instruction
function
Syntactic
instruction
Output format
Data feedback instances
1
Read the date
$01LT
$$0011LLTT
>00 2006/09/01 13:47:59
2
Read measurement
value
$01LV
$$0011LLVV
>00 1.97pH M 25.0C
3
Identify Relay status $01LR
$$0011LLRR
>00 RLY1 OFF, RLY2
OFF, WASH OFF
4
Read measurement
value and status
$01LA $$0011LLAA
>00
2006/09/01 13:48:36
1.97pH M 25.0C
RLY1 OFF, RLY2 OFF,
WASH OFF
5
Read the final
calibration data
$01LC $$0011LLCC
>00 CAL pH 0MV 100.0%
2004/01/01 00:00:00
6 Enter
set-up
mode
!01SI(2695) !!0011SSII(22669955)
>00
OK