40
Input of parameter settings in set-up mode
Items Instruction
function
Instruction
language Set the parameter range
1 Input the date
&01SD(xxxx)(yy)(zz)
xxxx: 2000~2099 year
yy : 1~12 month
zz : 1~31 day
2 Input the time
&01ST(xx)(yy)(zz)
xx: 0~23 hour
yy: 0~59 minute
zz: 0-59 second
3 Relay1
terminal
test
&01SH(x)
x: ON or OFF
4 Relay2
terminal
test
&01SL(x)
x: ON or OFF
5 WASH
terminal
test
&01SW(x)
x: ON or OFF
6 Set current pH/ORP way
&01RS(x)
x: 0 or 4 mA
7 Set current temperature way &01TS(x)
x: 0 or 4 mA
8 Set current pH/ORP value
&01RA(xxxx)(yyyy) Refer Note1
9 Set current temperature value &01TA(xxxx)(yyyy)
Refer Note2
10 WASH
time
setting
&01WT(xxxx)(yyyy)
xxxx: 0-9999 sec./ON
yyyy: 0-9999 H./Off
11 High point control setting &01HM(x)
x:
AUTO
or
OFF
12 Low point control setting
&01LM(x) x:
AUTO
or
OFF
13
Input PH/ORP high point
alarm value
&01HP(xxxx)(yyy)
Refer Note3
14
Input PH/ORP low point
alarm value
&01LP(xxxx)(yyy)
Refer Note4
15 Setting of measurement way &01MM(x)
x: pH or ORP
16
Setting of temperature
compensation way
&01TM(x)
x: AUTO/P
、
AUTO/N or
OFF
17
Setting of manual temperature
compensation value
Setting of automatic
temperature compensation
value
&01TV(xxxx)
Refer Note 5
18
Restore original factory
parameter settings
&01DS(ON)
19 Leave the set-up mode
&01SO