5. NEVER use positioning belts

a.   As a patient restraint.  A restraint requires a doctor’s

order.

b.   On a rider who is comatose or agitated.
c.   As a motor vehicle restraint.  In an accident or sudden

stop the rider may be thrown from the chair.
Wheelchair seat belts will not prevent this, and further
injury may result from the belts or straps.

If you fail to heed these warnings, damage to your chair, a fall, tip-
over or loss of control may occur and cause severe injury to the
rider or others.

J.   PUSH HANDLES

WARNING

1. Push handles provide secure points for an attendant to propel

and control the chair. This helps to prevent a fall or tip-over. 

2. Check to make sure push handle grips will not rotate or slip off. 

K. SEATING SYSTEMS

WARNING

1. Use of a seating system not approved by Sunrise may alter the

center of balance of this chair. This may cause a fall or tip-over. 

2.

Never

change the seating system of your chair unless you

consult your supplier first. 

3. Never raise your seating system to a height of more than

22" (measuring from the front of the seat pan to the floor)
with all actuators in their home position.

L.  UPHOLSTERY FABRIC

WARNING

1. Replace worn or torn fabric of the seat sling and seat back

as soon as you can. If you fail to do so, the seat may fail and
cause you to fall. Worn fabrics may increase the potential
for a fire hazard.

2. Sling fabric will deteriorate with age and use. Look for fray-

ing, thin spots, or stretching of fabric at rivet holes. Replace
fabric as required.

3. Be aware that washing may reduce flame retardation quali-

ties of the fabric.

M. SUSPENSION

WARNING

No component of the chair, was made with natural rubber latex,
Including the Suspension system.

N. WIRING

WARNING

Never pull on cables directly.  This can result in wires breaking
inside the connector or harness.  To remove a plug or connector,
always grasp the plug or connector itself. 

O. ASSIGNABLE CONTROLS

WARNING

It is possible to assign a function to buttons, or jacks of your wheel-
chair . If controls have been assigned to perform a dual or alternate
function ensure you know what each control operates. Please con-
tact your Authorize Dealer or the Quickie Technical Service
Department if you do not receive this information. 

P. POWER SEATING (SC,MPC)

WARNING

1.   Operation

• While driving this wheelchair in the reclined, elevated, or

tilted position use caution.

• Do Not attempt to operate any power seating option

while positioned on an incline.

• Do Not attempt to operate any power seating option

while this wheelchair is in motion.

• Do not attempt to operate the lift or tilt power seating

options around children.

• Through programming it is possible to reverse the direction

of all powered seating functions, ensure you know which
direction your seat is going to move before operating.

2.   Reduced Speed Driving (Creep)

This power chair is designed to automatically reduce the maximum
driving speed to a creep mode when certain power seating limits
are reached.

3.   Pinch points

Pinch points may occur when operating the powered seating
options on this chair. Ensure all hands and body parts are clear of
all power seating components that may result in pinch points
before operating them.

4.   Tilting Power Base with Push Handles

Do not attempt to tilt power wheelchairs by pulling down on the
push handles. Power wheelchairs have heavy bases. As such, attempt-
ing to tilt a power wheelchair to traverse obstacles can damage com-
ponents in the seating system and/or modular actuators.

5.   Remain Clear During Power Accessory Operation

Please stay clear of any powered actuator while the component is
in motion. Sunrise power mechanisms move over large ranges of
travel. Users should be aware of their surroundings while compo-
nents are in motion.

6.   Maximum User Weight

Do not exceed the weight capacity for your specific chair/seating sys-
tem configuration. Doing so could result in premature failure or injury.

7. Latch Mode

While operating any power seating function in latched mode use
caution. In latch mode, powered seating will not stop until a
reverse command is performed or the end of travel is met.

Q. S.P.O.T. 

(Single Power Option Tilt)

POWER ACCESSORY (

SC only

)

WARNING

The powered S.P.O.T. accessory for this chair has a maximum
user weight capacity of up to 300 lbs.  Never exceed this capacity
while operating the powered actuator.

WARNING

The powered S.P.O.T. should not be operated for more than
three minutes of continuous use.  Doing so may shorten the life of
the powered S.P.O.T.

VII. WARNINGS: COMPONENTS AND OPTIONS

Q UI C KI E   PU LS E

12

119832Rev. G

Summary of Contents for Pulse 5BC

Page 1: ...ble en formato PDF en nuestra p gina en Internet www SunriseMedical com Ingrese a la p gina del producto espec fico para descargar el manu al o comun quese con el proveedor autorizado de Sunrise Medic...

Page 2: ...ecyclable The separate collection and recycling of your product at the time of disposal will help conserve natural resources and ensure that it is disposed in a manner that protects the environment En...

Page 3: ...on 18 K Swing in Swing out Footrests 18 L Elevating Legrest 18 M Angle Adjustable Centermount Footrest 18 N Dual post Fixed height standard and Dual post Height Adjustable Armrests 19 O Single post He...

Page 4: ...to operate chair Standard Group 22 Battery Charger Off board 8 AMP Shroud Colors Red Black Blue Silver Copper Green Apple Yellow Pearl Pink Candy Purple Matte Black Midnight Blue Pearl white Digital c...

Page 5: ...see the icons listed below to identify warnings and potential haz ards WARNING These warnings must be followed failure to do so may cause personal injury or may dam age the wheelchair ESD Electrostat...

Page 6: ...C2 21 have been installed WARNING If there is any doubt that the specialty control input device is not immune to 20V m extra care should be taken around known sources of EMI WARNING There is no way to...

Page 7: ...afe use by your health care professional 2 Every wheelchair is different Take the time to learn the feel of this chair before you begin riding 3 Be aware that you must develop your own methods for the...

Page 8: ...gs axles motors or loosen fasteners J STREET USE WARNING 1 This product is not intended for street use Avoid streets whenever possible 2 Obey and follow all legal pedestrian pathways and laws that app...

Page 9: ...oystick and cause your chair to move when you do not expect it 2 Make sure motor brakes are engaged This keeps the chair from moving when you transfer 3 Work with your health care professional to lear...

Page 10: ...G Wheelchair lifts are used in vans buses and buildings to help you move from one level to another 1 Always turn off all power to your chair when you are on a lift If you fail to do so you may touch t...

Page 11: ...he seat Consult with the rider s doctor nurse or therapist to find out if the rider needs such a device 4 Use positioning belts only with a rider who can cooperate Make sure the rider can easily remov...

Page 12: ...Department if you do not receive this information P POWER SEATING SC MPC WARNING 1 Operation While driving this wheelchair in the reclined elevated or tilted position use caution Do Not attempt to op...

Page 13: ...occur WARNING To reduce the potential of injury to vehicle occupants wheelchair mounted accessories including but not limited to IV poles trays respiratory equipment backpacks and other personal items...

Page 14: ...rement points B Pelvic Belt Securement point C Attach WTORS to the tie down brackets in accordance with the manu facturer s instructions and SAE j2249 SECURING THE WHEELCHAIR This wheelchair is to be...

Page 15: ...air has been tested for seating in a motor vehicle with the factory installed seating system only Ensure that the factory installed seating system is secured to the wheelchair frame before operation R...

Page 16: ...8 ISO 10542 compliant rear secure ment straps be used to anchor the power chair during motor vehicle transport For power chairs that exceed the weight threshold mentioned above It is further rec ommen...

Page 17: ...chair before use 4 At least once a year have a complete safety check and service of your chair performed by a supplier B CLEANING 1 Paint Finish Clean the paint finish with mild soapy water monthly P...

Page 18: ...From the side place the pivot D into the top of the frame tube which guides the hanger into the correct alignment with the locking plate C on front frame tube b Rotate the hanger to the front until i...

Page 19: ...c Return lever to locked position d Push arm pad until upper armpost locks firmly into place 3 Removal a Press on the release lever and remove the armrest 4 Replacing Armrest a Slide armrest back int...

Page 20: ...Lock R BATTERY REMOVAL 1 To remove the batteries before transporting or servicing a Carefully flip back the seat see Section XI T Seat removal b Remove the shroud see Section XI E Shroud removal c Un...

Page 21: ...according to specifications turn your wheelchair OFF immediately and contact your authorized supplier to re program oper ational settings using the RNET or VR2 operating guide If you have any problems...

Page 22: ...e PG Drives Technology RNET control system B CONNECTORS Mating Connectors To connect the communication cables A Holding the connector housing firmly push the connector into its mate until you can no l...

Page 23: ...The available modes are dependent on programming and the range of auxil iary output devices connected to the control system Maximum Speed Profile Indicator This is a gauge that shows the maximum speed...

Page 24: ...Name A text string that displays the name of the currently selected profile Clock Displays the current time in a numeric format Color Joystick Speed Display Gives a proportional display of the wheelc...

Page 25: ...d If you operate the Joystick before or just after you switch the control system on the screen will flash the joystick displaced screen You must release and center the joystick to resume normal operat...

Page 26: ...yet left in a steady posi tion for a predetermined period of time The controller will auto matically turn off if the set time has been reached This feature is designed to protect the user batteries a...

Page 27: ...the batteries as soon as you can bars 1 7 If the battery gauge shows just red then you should charge the batteries immediately bars 1 3 3 On Off Button E Turning the chair on or off is accomplished by...

Page 28: ...only a certain number of times before they will fail and no longer hold a charge For answers to questions about batteries consult your Authorized Dealer 2 Use Proper Batteries Your chair operates on...

Page 29: ...r smoke or hold an open flame near batteries 2 Protect your skin and clothes a Always wear rubber gloves and safety glasses when you handle batteries 3 Corrosion Check battery terminals often for corr...

Page 30: ...d be set at a minimum of 60 in lbs 2 Dealer Service and Adjustment Reference materials Go to www sunrisemedical com for parts manuals instruction sheets and instructional videos that will aid in the r...

Page 31: ...Unplug the main power connect B d Lift the batteries out of the frame using the battery strap as a lifting handle 2 Battery installation See section X Batteries for installation WARNING Batteries may...

Page 32: ...rmrest until it starts to height adjust the armrest again c When you reach desired height remember to level the armrest using screws D and E as shown in step 2 above NOTE The cantilever armrest has ab...

Page 33: ...osen the clamp A move the joystick and then hand tighten the clamp handle b To move the joystick to the opposite side of the chair remove the hardware B and move the clamp and joystick receiver to the...

Page 34: ...or lowered b Using a 5mm hex key loosen the two screws F at the top of the footrest bracket and move up or down to set the correct height c Re tighten the hardware to hold the adjustment in place Q P...

Page 35: ...ed to replace modules or adjust the seat width of the chair a Remove the 4 screws E to detach the seat pan from the modules underneath The locations of the screws may vary depending on the size of the...

Page 36: ...ioning depending on your specific configuration Seat to floor heights and Pre tilt positioning should only be altered by a trained medical professional For the correct Pre tilt settings and seat to fl...

Page 37: ...Position 2 2 D Position 1 1 E and position 0 default F a Using a 5mm Hex key remove the M8 flat head screws 3 on each side b Position the backrest at the desired height The top screw should be mounte...

Page 38: ...roller a Flip up the armrest so that it is perpendicular to the floor b Loosen the thumbscrew E on the controller bracket c Slide the controller in or out of the armrest to the desired position d Tigh...

Page 39: ...r replace covered parts This is the exclusive remedy for consequential damages F WHAT YOU MUST DO 1 Obtain an authorized supplier while this warranty is in effect prior approval for return or repair o...

Page 40: ...per instructions in this manual to insure electromagnetic emissions are at acceptable levels TABLE 2 Immunity Test IEC 60601 test level Compliance Level Electromagnetic Environment Guidance Electrost...

Page 41: ...E 3 Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Portable and mobile RF communications equipment should be used no closer to any part of the product includi...

Page 42: ...o 80 MHz d 1 2 P 80 MHz to 80 MHz d 1 2 P 800 MHz to 2 5 GHz d 1 2 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 Recommended separation distances between portable...

Page 43: ...119832 Rev G 43 QUICKIE PULSE NOTES...

Page 44: ...2016 Sunrise Medical US LLC 7 16 119832 Rev G Customer Service 800 333 4000 www SunriseMedical com Sunrise Medical 2842 Business Park Ave Fresno CA 93727 USA In Canada 800 263 3390...