5. ASSEMBLY AND INSTALLATION OF TRI-STAR WALL

 

 

 

Your  shipping  carton  will  contain  a  lamp  head  and  flex  arm,  an  extension 
arm, Melamine covered wall board  (2

4” W  X 16” L  X 0.78” THK OR 610mm 

W  X  407mm  L  X      20mm      THK)  with  U-bracket  and  transformer  enclosure 
mounted  on  the  board,  and  four  (4)  plastic  fasteners  to  capture  the  power 
cord    on    the    underside    of    the    arm.    Surface  mounting  lag  bolts  are  not 
supplied.  These  should  be  purchased  to  meet  local  codes,  and  to  ensure 
they are appropriate for your particular mounting surface. 
 

     The length of the board has been chosen so that two wall studs may be   

engaged during installation. 

 
 
 

  WARNING! Please check local building codes for your specific application. 

 

 

ATTACHMENT OF LAMP HEAD/ARM/ASSEMBLY: 

1.  Attaching extension arm to wallboard: Place the end of the extension 

arm inside U-bracket and place plastic washers so that washers are 
between U-bracket and end of arm. Secure arm end to U-bracket with 
countersunk screws (2 X 1/4

”-20), use provided Allen key (5/32). (See 

following illustrations). 

 
 
 
 
 
 
 
 
 
 
 
 

 

END OF 

EXTENSION 

ARM 

Summary of Contents for TRI-STAR TS2015M

Page 1: ...SUNNEX MEDICAL LIGHT MOBILE TS2035M TS2015M WALL TS2035W TS2015W TRI STAR TM INSTALLATION AND SERVICE MANUAL...

Page 2: ...TALLATION OF TRISTAR WALL 5 6 ABOUT YOUR TRI STAR 7 7 CARE FOR YOUR TRI STAR LIGHT 8 8 SERVICE AND MAINTENANCE 9 9 WARRANTY 12 10 TROUBLESHOOTING 13 SAVE THIS PRODUCT MANUAL READ ALL INSTRUCTIONS BEFO...

Page 3: ...THE SHIPMENT REPORT SHORTAGES AND OR ERRORS TO SUNNEX OR YOUR DEALER S CUSTOMER SERVICE DEPARTMENT WITHIN FIVE 5 WORKING DAYS ALL FREIGHT CLAIMS BY EITHER THE DEALER OR CONSIGNEE SHOULD BE FILED WITH...

Page 4: ...n Bulb 12V 3x35W 10 20 W MODEL Halogen Bulb 12V 3x20W 10 Color temp 3 000K Medical Standard Class II Listing Weight 25 lbs Mobile 20 4 lbs Wall This product is rated as a Class II device mode of opera...

Page 5: ...LE Remove the upright column and the lamp head assembly from the shipping carton Insert the threaded end of the upright column and screw the upright column into the base Mount the screw with washer to...

Page 6: ...d These should be purchased to meet local codes and to ensure they are appropriate for your particular mounting surface The length of the board has been chosen so that two wall studs may be engaged du...

Page 7: ...4 Plug the lamp head cord into the outlet in the transformer enclosure 5 Plug the three prong power cord from the transformer enclosure into the wall outlet 6 ABOUT YOUR TRI STAR TRI STAR LAMP HEAD T...

Page 8: ...n allows you one hand control of the height of your lamp Simply squeeze the control at the top of the column and you can raise or lower the column easily The lamp head positioning arm connects with an...

Page 9: ...FLUIDS TO THE INSIDE OF THE LAMP HEAD HOUSING The metal components of your light are powder coated for long life They too can be cleaned by using a damp cloth of mild soap and water solution mineral...

Page 10: ...humbs and place two index fingers at flat part of spring two red lines in figure above push the spring away from you pulling it out from the slot Keep the other two ends of spring inside its holes 6 R...

Page 11: ...cifications USA 110V Buss Fuse MDA 2 250 Volt OR Interpower 813MDA 2 250V 2 Amp Fuse MUST USE A SLOW BLOW FUSE Europe 230V Littlefuse T 1 25A 218 250V 1 25 Amp Fuse MUST USE A SLOW BLOW FUSE 6 Follow...

Page 12: ...ICULAR NEEDS IF IMPLIED WARRANTIES MAY NOT BE DISCLAIMED UNDER APPLICABLE LAW THEN ANY IMPLIED WARRANTIES THAT MAY BE IMPOSED BY LAW ARE LIMITED IN DURATION TO THE LIMITED WARRANTY PERIOD SET FORTH HE...

Page 13: ...Sunnex receives the Product Standard shipments to the Customer will be paid by Sunnex Sunnex assumes no responsibility for shipment delays by the carrier f If Customer requires warranty service to be...

Page 14: ...e No power to the transformer box Check for facility power to the transformer box Transformer is malfunctioning Check for 14 V AC output at the transformer when it is under load Incorrect bulb install...

Page 15: ...www sunnexmedical com Sunnex Inc 8001 Tower Point Drive Charlotte NC 28227 Ph 800 445 7869 Fax 888 668 1920 REV 4 September 22 2019...