background image

10 

Sunnex TRI-STAR Service & Installation 
Manual 

 

 

 

 

 
 

 

 

 

Push Here 

To Move 

 

 
 
 
 
 
 
 
 
 
 

 
 
 
 

 
 
 
CAUTION: ONLY USE DICHROIC BULBS; ANY DEVIATION FROM THE DIRECTIONS 
MAY SERIOUSLY DAMAGE YOUR LIGHT. 35W IS MAXIMUM ALLOWED.

 

 

When a lamp/reflector burns out, ALLOW SUFFICIENT TIME FOR IT TO COOL 
BEFORE ATTEMPTING A REPLACEMENT. Once cool, you should: 

 

1.  Refer to figure above. Turn the lamp head over so the three glass 

assemblies face upward. 

2.  With one hand supporting the backside of the lamp, grasp the notched 

ring (1) and twist it counter clockwise. 

3.  Grasp the bulb with one hand and unplug socket connector with the other. 
4.  The notched ring assembly can now be removed from the lamp head and placed 

face down on a flat surface. 

5.  Hold the notched ring assembly with thumbs and place two index 

fingers at flat part of spring (two red lines in figure above), push the spring away 
from you pulling it out from the slot (Keep the other two ends of spring inside its 
holes) 

6.  Remove old bulb 
7.  Install new bulb. Do not touch inside reflector surface of bulb, always grasp 

from bulb outside edges 

8.  Close the retainer spring on the bulb by inserting the front tab of 

spring in the slot, so that it fits properly. Check to ensure proper seating of the 
spring, by pulling on the bulb. [Position (1) in figure above i

s “CLOSED 

SPRING

”]. Ensure that two open ends of spring are properly seated into holes in 

the notched ring assembly. 

9.  Reconnect plug socket to bulb. 
10.  Replace the notched ring assembly in the lamp head by aligning the 

three guide projections into the slots on the lamp head. Press gently inward as you 
twist the assembly clockwise. You will hit a stop and set in place. 

11.  The light is now ready for use. 
 

 
 
 

Summary of Contents for TRI-STAR TS2015M

Page 1: ...SUNNEX MEDICAL LIGHT MOBILE TS2035M TS2015M WALL TS2035W TS2015W TRI STAR TM INSTALLATION AND SERVICE MANUAL...

Page 2: ...TALLATION OF TRISTAR WALL 5 6 ABOUT YOUR TRI STAR 7 7 CARE FOR YOUR TRI STAR LIGHT 8 8 SERVICE AND MAINTENANCE 9 9 WARRANTY 12 10 TROUBLESHOOTING 13 SAVE THIS PRODUCT MANUAL READ ALL INSTRUCTIONS BEFO...

Page 3: ...THE SHIPMENT REPORT SHORTAGES AND OR ERRORS TO SUNNEX OR YOUR DEALER S CUSTOMER SERVICE DEPARTMENT WITHIN FIVE 5 WORKING DAYS ALL FREIGHT CLAIMS BY EITHER THE DEALER OR CONSIGNEE SHOULD BE FILED WITH...

Page 4: ...n Bulb 12V 3x35W 10 20 W MODEL Halogen Bulb 12V 3x20W 10 Color temp 3 000K Medical Standard Class II Listing Weight 25 lbs Mobile 20 4 lbs Wall This product is rated as a Class II device mode of opera...

Page 5: ...LE Remove the upright column and the lamp head assembly from the shipping carton Insert the threaded end of the upright column and screw the upright column into the base Mount the screw with washer to...

Page 6: ...d These should be purchased to meet local codes and to ensure they are appropriate for your particular mounting surface The length of the board has been chosen so that two wall studs may be engaged du...

Page 7: ...4 Plug the lamp head cord into the outlet in the transformer enclosure 5 Plug the three prong power cord from the transformer enclosure into the wall outlet 6 ABOUT YOUR TRI STAR TRI STAR LAMP HEAD T...

Page 8: ...n allows you one hand control of the height of your lamp Simply squeeze the control at the top of the column and you can raise or lower the column easily The lamp head positioning arm connects with an...

Page 9: ...FLUIDS TO THE INSIDE OF THE LAMP HEAD HOUSING The metal components of your light are powder coated for long life They too can be cleaned by using a damp cloth of mild soap and water solution mineral...

Page 10: ...humbs and place two index fingers at flat part of spring two red lines in figure above push the spring away from you pulling it out from the slot Keep the other two ends of spring inside its holes 6 R...

Page 11: ...cifications USA 110V Buss Fuse MDA 2 250 Volt OR Interpower 813MDA 2 250V 2 Amp Fuse MUST USE A SLOW BLOW FUSE Europe 230V Littlefuse T 1 25A 218 250V 1 25 Amp Fuse MUST USE A SLOW BLOW FUSE 6 Follow...

Page 12: ...ICULAR NEEDS IF IMPLIED WARRANTIES MAY NOT BE DISCLAIMED UNDER APPLICABLE LAW THEN ANY IMPLIED WARRANTIES THAT MAY BE IMPOSED BY LAW ARE LIMITED IN DURATION TO THE LIMITED WARRANTY PERIOD SET FORTH HE...

Page 13: ...Sunnex receives the Product Standard shipments to the Customer will be paid by Sunnex Sunnex assumes no responsibility for shipment delays by the carrier f If Customer requires warranty service to be...

Page 14: ...e No power to the transformer box Check for facility power to the transformer box Transformer is malfunctioning Check for 14 V AC output at the transformer when it is under load Incorrect bulb install...

Page 15: ...www sunnexmedical com Sunnex Inc 8001 Tower Point Drive Charlotte NC 28227 Ph 800 445 7869 Fax 888 668 1920 REV 4 September 22 2019...

Reviews: