background image

11 

Sunnex TRI-STAR Service & Installation 
Manual 

 

 

 

 

Only use the following bulbs: 

Sunnex part# 18562

 

 

 

CAUTION!  A CHOICE OF  A REPLACEMENT LAMP/REFLECTOR CAN AFFECT THE 
PERFORMANCE OF YOUR TRISTAR LIGHT. 

  

 

8.2 FUSE REPLACEMENT

 

 

 

1.  Remove the plug from wall receptacle. 

 

 

2.  Access the transformer housing and detach the front cover by removing 

the four plastic screws using a flat screwdriver. 

3.  Locate fuse blocks (2) in the transformer. 

 

 

4.  Turn the fuse holder counter clockwise using a suitable tool and remove 

the fuse holder from the fuse block by pulling it out. 

5.  Replace fuse with proper replacement fuse. 

Fuse Replacement Specifications: 

USA (110V): Buss Fuse 

– MDA 2 250 Volt OR Interpower 813MDA-2 

(250V, 2 Amp Fuse,

 MUST USE A SLOW-BLOW FUSE

) Europe 

(230V): Littlefuse 

– T-1.25A-218 (250V, 1.25 Amp Fuse, 

MUST USE A SLOW-BLOW FUSE

6.  Follow procedure in reverse to return lamp to operating condition. 

 

 

 

For more information on our product, visit our web site at 

www.sunnexonline.com 

or call us at (800) 445-7869. 

 

SYMBOLS

 

 

 

 

 

 

 

 

 

 

 

GROUND

 

 

SWITCH ON

 

 

SWITCH OFF

 

 

CAUTION

 

 
 

9. WARRANTY

 

Limited Warranty 
Sunnex warrants that the Product, if properly used, will be free from defects in material and 
workmanship, for a period of three (3) years from the date of shipment to Customer.  This Limited 
Warranty extends only to the original buyer. If the Product is determined to be defective within 
the warranty period, Sunnex, at its option, will repair or replace the Product, at no charge. 

 
 
 

Summary of Contents for TRI-STAR TS2015M

Page 1: ...SUNNEX MEDICAL LIGHT MOBILE TS2035M TS2015M WALL TS2035W TS2015W TRI STAR TM INSTALLATION AND SERVICE MANUAL...

Page 2: ...TALLATION OF TRISTAR WALL 5 6 ABOUT YOUR TRI STAR 7 7 CARE FOR YOUR TRI STAR LIGHT 8 8 SERVICE AND MAINTENANCE 9 9 WARRANTY 12 10 TROUBLESHOOTING 13 SAVE THIS PRODUCT MANUAL READ ALL INSTRUCTIONS BEFO...

Page 3: ...THE SHIPMENT REPORT SHORTAGES AND OR ERRORS TO SUNNEX OR YOUR DEALER S CUSTOMER SERVICE DEPARTMENT WITHIN FIVE 5 WORKING DAYS ALL FREIGHT CLAIMS BY EITHER THE DEALER OR CONSIGNEE SHOULD BE FILED WITH...

Page 4: ...n Bulb 12V 3x35W 10 20 W MODEL Halogen Bulb 12V 3x20W 10 Color temp 3 000K Medical Standard Class II Listing Weight 25 lbs Mobile 20 4 lbs Wall This product is rated as a Class II device mode of opera...

Page 5: ...LE Remove the upright column and the lamp head assembly from the shipping carton Insert the threaded end of the upright column and screw the upright column into the base Mount the screw with washer to...

Page 6: ...d These should be purchased to meet local codes and to ensure they are appropriate for your particular mounting surface The length of the board has been chosen so that two wall studs may be engaged du...

Page 7: ...4 Plug the lamp head cord into the outlet in the transformer enclosure 5 Plug the three prong power cord from the transformer enclosure into the wall outlet 6 ABOUT YOUR TRI STAR TRI STAR LAMP HEAD T...

Page 8: ...n allows you one hand control of the height of your lamp Simply squeeze the control at the top of the column and you can raise or lower the column easily The lamp head positioning arm connects with an...

Page 9: ...FLUIDS TO THE INSIDE OF THE LAMP HEAD HOUSING The metal components of your light are powder coated for long life They too can be cleaned by using a damp cloth of mild soap and water solution mineral...

Page 10: ...humbs and place two index fingers at flat part of spring two red lines in figure above push the spring away from you pulling it out from the slot Keep the other two ends of spring inside its holes 6 R...

Page 11: ...cifications USA 110V Buss Fuse MDA 2 250 Volt OR Interpower 813MDA 2 250V 2 Amp Fuse MUST USE A SLOW BLOW FUSE Europe 230V Littlefuse T 1 25A 218 250V 1 25 Amp Fuse MUST USE A SLOW BLOW FUSE 6 Follow...

Page 12: ...ICULAR NEEDS IF IMPLIED WARRANTIES MAY NOT BE DISCLAIMED UNDER APPLICABLE LAW THEN ANY IMPLIED WARRANTIES THAT MAY BE IMPOSED BY LAW ARE LIMITED IN DURATION TO THE LIMITED WARRANTY PERIOD SET FORTH HE...

Page 13: ...Sunnex receives the Product Standard shipments to the Customer will be paid by Sunnex Sunnex assumes no responsibility for shipment delays by the carrier f If Customer requires warranty service to be...

Page 14: ...e No power to the transformer box Check for facility power to the transformer box Transformer is malfunctioning Check for 14 V AC output at the transformer when it is under load Incorrect bulb install...

Page 15: ...www sunnexmedical com Sunnex Inc 8001 Tower Point Drive Charlotte NC 28227 Ph 800 445 7869 Fax 888 668 1920 REV 4 September 22 2019...

Reviews: