Troubleshooting
Problem
Possible Cause
Remedy
Warning indicator on with
unit in Standby mode.
Unit shut down in an over temperature
condition.
Empty the reservoir and refill with room
temperature water.
Make sure all clamps are open.
Press the
On/Standby
button.
Verify flow through the pad.
The
Warning
light will turn off within 5
minutes.
Flow indicator and Standby
indicator are on with
T/Pump not pumping.
Unit detected a Flow warning for more
than 5 minutes, thus goes to standby.
Reference “Flow indicator light is on”
above.
Correct the problem, and press the
On/
Standby
to put the unit back into Run
mode.
Temperature Setpoint light
blinking.
Unit is warming up to the selected
setpoint.
Unit is in Cooling mode, for longer than
40 minutes.
Indication only.
Follow the shutdown procedure. Drain the
water in reservoir to ice fill level, and refill
with ice. Follow the start-up procedure.