Control Panel and Display
9
Features
Convenient Preparation
♦
The internal battery is charged automatically whenever the SmartPump is plugged into
an AC power source. The internal battery is recharged to approximately 80% level within
one hour.
♦
The fully charged internal battery provides five hours of uninterrupted power. However,
internal battery power is a safety feature and is to be used as a back up only. Always
connect the SmartPump to an AC power source for normal operation.
♦
Battery ‘saver’ technology keeps the internal battery charged. If the SmartPump is fully
charged, then left ‘OFF(standby)’ and unplugged, the internal battery is kept charged for
up to 30 days.
♦
The SmartPump is center mounted to its recommended Stryker roll stand with pole.
♦
Fill lines are attached to the positive locking connectors located at the bottom corners of
the SmartPump.
♦
At start-up, the SmartPump performs a self-test automatically and tests the pneumatic,
pressure, display, battery, and processor systems. The SmartPump will alert the user via
alarms, indicators and the display of error codes to assist in troubleshooting.
Ease of Operation
♦
The procedure timer and cuff target pressure may be set using simple, intuitive steps.
Control buttons are clearly identified.
♦
Inflation is initiated by pressing the
Inflate
button.
♦
Deflation is initiated by pressing and holding the
Deflate
button for 1.5 seconds.
Continuous, Real-time Monitoring
♦
The large, bright, backlit LCD display shows total elapsed cuff inflation time and cuff
pressure (mmHg). The displayed time format is user selectable: minutes or hours and
minutes.
♦
Audible control alarms and flashing graphical symbols identify conditions requiring
attention.
Accessory Information*
Description REF
Roll Stand with Pole …..…….………………………….…………..5920-013-000
Tubing, Stockinettes, Fill Lines, and Adapters .………………..5920-xxx-xxx Series
Disposable Tourniquet Cuffs ………………………..….…………5921-xxx-xxx Series
Non-sterile Reusable Tourniquet Cuffs…..…………….………..5922-xxx-xxx Series
*Contact your Stryker sales representative for a complete list of accessories.