General Use Procedure
28
Cuff Inflation
The individual channels for Cuff 1 and Cuff 2 operate independently. The
SmartPump display screen displays real-time pressure values and total
elapsed cuff inflation time for each cuff.
Intravenous Regional Anesthesia (IVRA) Lock
The Intravenous Regional Anesthesia (IVRA) Lock may be activated either
prior to or after inflation. The lock prevents accidental deflation of an
active cuff associated with a Bier Block procedure or the management of
two cuffs involved in a bi-lateral procedure.
In its unlocked state, the IVRA Lock icon resembles an unlocked padlock.
In a locked state, the icon changes to a locked padlock. See images
below:
The IVRA (
lock, unlock
) button is below the LCD, on the right.
If, when attempting to deflate the second cuff with the IVRA Lock enabled,
the SmartPump will produce an audible error-beep and the IVRA padlock
will blink.
To Inflate a Cuff
Press the
Inflate
button on the selected side.
♦
The selected cuff is inflated to the target pressure.
♦
The SmartPump begins tracking and displaying the elapsed cuff time.
The elapsed time appears at the top of the screen and is represented
numerically and graphically.
♦
The cuff pressure is monitored on a real-time basis. The pressure is
displayed numerically at the bottom of the screen.
♦
The graphic pressure gauge appears above the numeric pressure and
graphically represents the status of the cuff.
IVRA
Lock Button