Specifications 48
7 Specifications
Air Pump:
Nominal power:
12 VDC, 1.25 A
Flow:
10 LPM open flow
Pressure Transducers:
Operating pressure:
0 – 15 psi
Maximum pressure:
30 psi
Repeatability:
± 0.2% FS typical
Cuff Pressure:
Accuracy:
± 2% or 4 mmHg, whichever is greater, set pressure value
Range:
100 to 475 mmHg
Cuff Timer:
Range:
1 to 240 minutes
Adjustment:
1 to 15 minutes – 1 minute increments;
15 to 240 minutes – 5 minute increments
Quick Connect Hose
Couplings
CPC PMC1602 or equivalent
Environment
Operation Storage
and
Transportation
Temperature:
Relative Humidity:
Atmospheric Pressure:
Optional Printer Tray:
Tray size:
8.25 inch width x 7.5 inch depth [21 cm x 19 cm]
Tray load capacity:
5 lbs [2.3 kgs] maximum load capacity
Pole, Basket and Stand:
Pole:
adjustable height from 46 inch to 56 inch [117cm to 142cm]
Base:
weighted base with 5 legs and heavy duty casters (two casters can be
locked)
Basket size:
6 inch height x 13 inch width x 8 inch depth [15.2cm x 33.0cm x 20.3cm]
Basket load capacity:
5 lbs [2.3 kgs] maximum load capacity