At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may need
to perform preventive maintenance checks more often based on your level of product usage.
Remove product from service before performing preventive maintenance.
N
No
otte
e -- Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items for the standard models:
Zipper and covers (top and bottom) are free of tears, cuts, holes, or other openings
Internal components for signs of staining from fluid ingress or contamination by fully unzipping the covers
Labels for legibility, proper adherence, and integrity
Handles are free of rips or cracks and stitching is intact
Foam has not degraded or come apart
Fire barrier for rips, cracks, or other visible signs of damage
Inspect the following items for the behavioral health models:
Covers (top and bottom) are free of tears, cuts, holes, or other openings
Labels for legibility, proper adherence, and integrity
Product serial number:
Completed by:
Date:
C
Clle
ea
an
niin
ng
g a
an
nd
d d
diissiin
nffe
eccttiin
ng
g w
wiitth
h w
wiip
pe
ess
For United States only. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327-
0770.
Stryker’s preferred wipes (2060-000-001 6'' x 10'' or 2060-000-002 9'' x 12'') include the following active ingredients:
•
n-Alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium chloride - 0.154%
•
n-Alkyl (68% C12, 32% C14) dimethyl ethylbenzyl ammonium chloride - 0.154%
•
Isopropanol - 21.000%
Non-active ingredient: Ethylene Glycol Monobutyl Ether – < 3%
N
No
otte
e -- For safety information, read the product label.
To clean or disinfect the external product surface:
1. To clean, wipe external surfaces with a fresh, clean wipe to remove all visible soils. Repeat as necessary until the
product is clean.
N
No
otte
e
•
Use as many wipes as necessary.
•
Complete step 1 before you disinfect.
2. To disinfect, wipe external surfaces with a fresh, clean wipe until wet. Allow the external surface to remain wet for two
minutes at room temperature.
3. Allow the product to dry before you return it to service.
EN
10
2815-009-001 Rev D.1
Summary of Contents for ProForm 281505550007
Page 2: ......
Page 18: ......
Page 34: ......
Page 50: ......
Page 52: ...파손 주의 포장 개봉 시 날카로운 물체를 사용하지 말 것 장비 질량 KO 2815 009 001 Rev D 1 ...
Page 59: ...일 일련 련번 번호 호 위 위치 치 A 제 제조 조일 일 일련번호의 첫 네 자리가 제조 연도입니다 2815 009 001 Rev D 1 7 KO ...
Page 66: ......
Page 81: ......
Page 82: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 2815 009 001 Rev D 1 WCR 2019 04 ...