Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
C
CA
AU
UT
TIIO
ON
N
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the
user or patient or damage to the product or other property. This includes special care necessary for the safe and effective
use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse.
N
No
otte
e -- Provides special information to make maintenance easier or important instructions clearer.
S
Su
um
mm
ma
arryy o
off ssa
affe
ettyy p
prre
ecca
au
uttiio
on
nss
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
W
WA
AR
RN
NIIN
NG
G
•
Always plug the power cord into the bed auxiliary outlet or a grounded power respectable. Do not use an extension cord.
•
The IIsso
oA
Aiirr system may interfere with patient ECG measurements. You may need to disconnect the AST cable, turn off
the IIsso
oA
Aiirr system, or move the patient to a non-powered system.
•
The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
•
Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be no closer than 12 inches (30 cm) to any part of the IIsso
oA
Aiirr system, that includes cables specified by the manufacturer.
•
Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the products. If
such use is necessary, carefully observe stacked or adjacent equipment to make sure that they are operating properly.
•
This equipment/system is intended for use by healthcare professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such
as reorienting or relocating IIsso
oA
Aiirr system or shielding the location.
C
CA
AU
UT
TIIO
ON
N
•
Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
•
Always use ESD protective equipment before you open antistatic bags and service electronic parts.
•
Do not place unprotected circuit boards on the floor.