2. Using a 3 mm hex driver, remove the six bolts (F) and nuts (G) that secure the six solenoid valves (D) to the pump
manifold assembly (Figure 28). Each nut holds the pump manifold tight to keep the pump manifold from turning. Save
the bolts and nuts.
D
E
F
G
G
F
Fiig
gu
urre
e 2
28
8 –– S
So
olle
en
no
oiid
d vva
allvve
e
N
No
otte
e -- When you reinstall the bolts (G), tighten the bolts down in an X pattern.
3. Remove and save the solenoid retaining plate (E) from the pump manifold assembly.
4. Unscrew and remove the solenoid valve (D) to be replaced.
N
No
otte
e -- There are six solenoid valves. Five of the six solenoid valves are identical. The bottom right solenoid valve is for
LAL and has a white cap for identification.
5. Plug the pump cord into an accessible outlet.
6. Run the pump diagnostic test. All tests must pass.
7. Reverse steps to reinstall.
8. Verify proper operation before you return the product to service.
P
PC
CB
BA
A a
asssse
em
mb
bllyy rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
#2 Phillips screwdriver
•
ESD system
EN