13. Discard the pump connector and four hoses.
14. Reverse steps to reinstall.
15. Plug the pump cord into an accessible outlet.
16. Run the pump diagnostic test. All tests must pass.
N
No
otte
e -- Watch each test when you run the pump diagnostics. Confirm that the correct cell inflates the air hoses AS, BS,
SBR, and LAL. The diagnostic test cannot identify if the connected lines are correct or incorrect.
A
Aiirr
h
ho
osse
ess
D
De
essccrriip
pttiio
on
n
AS
A cell line
BS
B cell line
SBR
Side bolster line
LAL
LAL line
17. Verify proper operation before you return the product to service.
S
Su
up
pp
po
orrtt ssu
urrffa
acce
e tto
op
p cco
ovve
err rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
None
P
Prro
occe
ed
du
urre
e::
1. Raise the bed height to the highest height position.
2. Place the litter surface in a flat, horizontal position.
3. Unzip the top cover.
4. Remove and discard the top cover.
5. Attach the new top cover. Start the zipper at the head end and stop at the foot end.
6. Verify proper operation before you return the product to service.