1. See
Separate the front and back covers
C
CA
AU
UT
TIIO
ON
N
•
Always use ESD protective equipment before you open antistatic bags and service electronic parts.
•
Do not place unprotected circuit boards on the floor.
2. Unplug the inlet power cord assembly from the power supply CN1 (two wire cable).
3. Using a #2 Phillips screwdriver, remove the four screws (S) that secure the power supply (N) to the main support plate
(R) (Figure 24). Save the screws.
R
S
N
F
Fiig
gu
urre
e 2
24
4 –– P
Po
ow
we
err ssu
up
pp
pllyy tto
o m
ma
aiin
n
4. Remove and discard the power supply (N) (Figure 24).
N
No
otte
e -- Do not dispose of as unsorted municipal waste. Refer to your local distributor for return or collection systems
available in your country.
5. Reverse steps to install the new power supply.
6. Reconnect the support surface.
7. Run a diagnostics and full function test before you return the product to service.