F
Fiig
gu
urre
e 6
6 –– N
No
on
n--A
AS
ST
T a
aiirr cce
ellll rre
ep
plla
acce
em
me
en
ntt
10. Unbutton both sides of the non-AST air cell from the side bolsters.
11. Slide the non-AST air cell out of the center belt to remove.
12. Remove and discard the non-AST air cell.
13. Insert the new non-AST air cell through the center belt.
14. Using rubbing alcohol, lubricate the two line manifold fitting and the non-AST air cell fitting.
15. Apply firm, constant force while you turn the air cell fitting back and forth to connect the non-AST air cell to the two line
manifold.
16. Install a nylon cable tie onto the two line manifold/non-AST air cell fitting, 0.25 inches from the end of the hose.
17. Inflate and verify that the air cell inflates.
18. Button both sides of the non-AST air cell to the side bolsters.
19. Move the fire barrier over the air cells toward the foot end to install the fire barrier.
20. Clip the foot end center belt retainer to the bottom cover.
21. Zip the fire barrier.
22. Zip the support surface top cover onto the bottom cover.
23. Run the pump diagnostic test. All tests must pass.
24. Verify proper operation before you return the product to service.
A
AS
ST
T ffo
ou
urr cce
ellll kkiitt rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Diagonal pliers
•
Rubbing alcohol
P
Prro
occe
ed
du
urre
e::
1. Raise the bed height to the highest position.
2. Place the litter surface in a flat, horizontal position.
2941-109-002 Rev AA.0
13
EN