280C-009-002 REV B
1-17
English
Preventative Maintenance
Procedure:
Hi-Lo Actuators
1 . Lower the bed to the lowest position and apply the brakes . Flatten the litter .
2 . Lower the four siderails .
3 . Using a #2 Phillips screwdriver, remove the head and foot section plastic covers (if equipped), to reach the Hi-Lo
actuators, or raise the head section to the highest position and fold the foot section back towards the head end of
the bed . Secure the foot section using a bungee cord .
4 . Unplug the power cord from the wall receptacle .
5 . Using a brush, apply OG2 grease to the screw threads of both actuators through the openings on the sides of the
dust tubes . Ensure the grease penetrates to the bottom of the threads .
6 . Lower the foot section .
7 . Plug the power cord into the wall receptacle .
8 . Using the electronic controls, raise and lower the bed several times to spread the grease evenly .
ACTUATOR SCREW LUBRICATION (CONTINUED)