1-164
280C-009-002 REV B
English
The GoBed
®
II MedSurg Bed, Model FL28C is suited for use in the electromagnetic environment specified below .
The customer or the user of the GoBed
®
II MedSurg Bed, Model FL28C should assure that it is used in such an
environment .
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2 .5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the GoBed
®
II MedSurg Bed, Model FL28C,
including cables, than the recommended
separation distance calculated from the
equation appropriate for the frequency of the
transmitter .
Recommended Separation Distance
d=1 .2
d=1 .2
d=2 .3
800 MHz to 2 .5 GHz
where
P
is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in meters
(m) .
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range .
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies .
Note 2
These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people .
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy .
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered . If the measured field strength in the location in which the GoBed
®
II MedSurg Bed, Model FL28C is used
exceeds the applicable RF compliance level above, the GoBed
®
II MedSurg Bed, Model FL28C should be observed
to verify normal operation . If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the GoBed
®
II MedSurg Bed, Model FL28C .
b
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m .
GOBED
®
II MEDSURG BED, MODEL FL28C (CONTINUED)
EMC Information
Summary of Contents for FL28C
Page 2: ......
Page 83: ...www stryker com 280C 009 002 REV B 1 81 Return To Table of Contents English Notes ...
Page 94: ...1 92 280C 009 002 REV B www stryker com Return To Table of Contents English Notes ...
Page 111: ...www stryker com 280C 009 002 REV B 1 109 Return To Table of Contents English Notes ...
Page 170: ......
Page 251: ...www stryker com 280C 009 002 REV B 2 81 Retour à la table des matières Français Commentaires ...
Page 262: ...2 92 280C 009 002 REV B www stryker com Retour à la table des matières Français Commentaires ...
Page 279: ...www stryker com 280C 009 002 REV B 2 109 Retour à la table des matières Français Commentaires ...
Page 336: ......
Page 337: ......