1-52
280C-009-002 REV B
English
TRANSFORMER REPLACEMENT (100/200/220/240 VAC OPTIONS)
Tools Required:
•
#2 Phillips Screwdriver
•
1/2” Combination Wrench
Procedure:
1 . Raise the bed to the highest position and apply the brakes .
2 . Unplug the power cord from the wall receptacle .
3 . Using a #2 Phillips screwdriver, remove the six screws (A) holding the cover to the litter . Keep the cable clips . See
Figure 12,
4 . Properly ground yourself (see “Grounding Diagram” section,
) .
5 . Using a #2 Phillips screwdriver, remove the nut, washer and bolt (A) holding the transformer to the litter . Disconnect
the transformer cable and remove it .
Note
Use caution to avoid overtightening the bolt .
6 . Reverse the above steps to install the new transformer .
7 . Test the bed for proper operation before returning the bed to service .