Maintenance Manual
24
5. While supporting the rail, remove the two last
screws and the bungee cord.
6. Lift the siderail, pass it over the mechanism
arms and lay it on the mattress support while
gently pulling on the cables (see opposite
illustration).
7. Using a Phillips screwdriver, remove the six
screws (I) holding the aluminum structure to
the siderail.
8. Grasp and lift the upper part of the aluminum
structure lying on the mattress support until it
may be secure temporarily to one of the
mechanism arms using a cable tie.
9. Remove from the defective board the cable connected to it. The optional nurse call cable
may also be present, remove it.
10. Using a Phillips screwdriver, remove the eight screws (J) holding the control board to the
siderail.
11. Reverse the above steps to install the new control board.
12. Test all controls of the inner/outer control panels for proper operation before reassembling
the siderail and returning the bed to service.
SIDERAIL INNER PANEL SPEAKERPHONE REPLACEMENT
Required Tools:
Small Slotted Head Screwdriver
Phillips Screwdriver
Side Cutters
Bungee Cord
3/16" Allen Key
Soldering Iron and Soft Welding Wire
Procedure:
1. Follow steps 1 to 8 of the siderail inner panel control board replacement procedure,
described above.
2. Using a Phillips screwdriver, remove the four screws (K, fig. 4.1C, page 22) holding the
speakerphone to the siderail.
3. Using a soldering iron, unsolder the two wires attached to the defective speakerphone and
remove the speakerphone. Do the same with the wires attached to the new speakerphone.
Keep the wires as replacement parts.
4. Using a soldering iron and soft welding wire, attach the two wires to the new speakerphone.
Clean the welds.
N
N
O
O
T
T
E
E
Before reassembling the different parts of the siderail, we recommend that you check the
continuity at the speakerphone terminals to ensure the circuit operates properly.
5. Reverse the preceding steps to install the new speakerphone.
6. Test the nurse call and all controls of the inner/outer control panels for proper operation
before reassembling the siderail and returning the bed to service.