![Stryker EPIC II Maintenance Manual Download Page 7](http://html.mh-extra.com/html/stryker/epic-ii/epic-ii_maintenance-manual_1385890007.webp)
Return To Table of Contents
www.stryker.com
2030-809-002 REV D
7
INTENDED USE
This manual is designed to assist you with the
maintenance
of Stryker Model 2030 Epic II
®
Critical Care Bed. Carefully
read this manual thoroughly before using the equipment or beginning maintenance on it. To ensure safe operation of
this equipment, it is recommended that methods and procedures be established for educating and training staff on the
safe operation of this bed.
Note
Epic II
®
PLUS and Epic II
®
+ are interchangeable terms within this manual.
Introduction
SPECIFICATIONS
Safe Working Load
Note:
Safe Working Load indicates the sum of the
patient, mattress, and accessory weight.
500 lbs
227 kg
Bed Weight
594 lbs
269 kg
Scale System Capacity (optional equipment). Loads weighing up to
500 lbs
227 kg
Scale System Accuracy (optional equipment)
±1 pound of total patient weight at any bed
position
1
(patients weighing 100 pounds or less)
±1% of total patient weight at any bed position
1
(patients weighing greater than 100 pounds)
Overall Length/Width
Length
91”
231 cm
Width
42.5”
108 cm
Minimum Bed Height
18”
46 cm
Maximum Bed Height (Standard)
Maximum Bed Height (Enhanced)
30” to 31”
32” to 33”
76.2 cm to 78.7 cm.
81.2 cm to 83.8 cm.
Fluoroscopy Access
Epic II®
17”
43.1 cm
Epic II® PLUS
16”
41 cm
Knee Gatch Angle
0° to 30°
Back Angle
0° to 90°
Trendelenburg/Reverse Trendelenburg
-12° to +12° ±2°
Electrical Requirements
120VAC, 60 Hz, 8.0 A
100VAC, 50/60 Hz, 9.0 A (Japan Option)
Battery Voltage
(2) 12V, 31 Ah (Gel Cell Battery) Zoom® Option
(1) 9V (Alkaline Battery) Nurse Call Option
Noise Level
> 65 dB
Outlet Option
120VAC, 5A, 60 Hz
1
If the bed is equipped with the enhanced height option, the scale accuracy is as described above for litter angles
from 0° to ±5° Trend.
MATTRESS SPECIFICATIONS
Thickness
6”
15.2 cm
Width
>= 35”
>= 88.9 cm
Length
>= 84”
>= 213.4
ILD
80 lbs
36.3 kg
The above stated mattress specifications assist in ensuring the product conforms to HBSW and IEC
specifications.
Stryker reserves the right to change specifications without notice.
Specifications listed are approximate and may vary slightly from unit to unit or by power supply fluctuations.
Summary of Contents for EPIC II
Page 2: ......
Page 71: ...Return To Table of Contents www stryker com 2030 809 002 REV D 71 Notes ...
Page 77: ...Return To Table of Contents www stryker com 2030 809 002 REV D 77 Base Assembly and Options ...
Page 78: ...Return To Table of Contents 78 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 80: ...Return To Table of Contents 80 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 96: ...Return To Table of Contents 96 2030 809 002 REV D www stryker com Notes ...
Page 106: ...Return To Table of Contents 106 2030 809 002 REV D www stryker com Litter Assembly ...
Page 107: ...Return To Table of Contents www stryker com 2030 809 002 REV D 107 Litter Assembly ...
Page 108: ...Return To Table of Contents 108 2030 809 002 REV D www stryker com Litter Assembly ...
Page 109: ...Return To Table of Contents www stryker com 2030 809 002 REV D 109 Litter Assembly ...
Page 110: ...Return To Table of Contents 110 2030 809 002 REV D www stryker com Litter Assembly ...
Page 111: ...Return To Table of Contents www stryker com 2030 809 002 REV D 111 Litter Assembly ...
Page 112: ...Return To Table of Contents 112 2030 809 002 REV D www stryker com Litter Assembly ...
Page 113: ...Return To Table of Contents www stryker com 2030 809 002 REV D 113 Litter Assembly ...
Page 120: ...Return To Table of Contents 120 2030 809 002 REV D www stryker com Notes ...
Page 171: ...Return To Table of Contents www stryker com 2030 809 002 REV D 171 Foot End Siderail Assembly ...
Page 226: ......
Page 227: ......