![Stryker EPIC II Maintenance Manual Download Page 42](http://html.mh-extra.com/html/stryker/epic-ii/epic-ii_maintenance-manual_1385890042.webp)
Return To Table of Contents
42
2030-809-002 REV D
www.stryker.com
LIFT HOUSING REMOVAL AND REPLACEMENT
A
FOOT END − BOTTOM VIEW
Tools Required:
•
#2 Phillips Screwdriver
•
Bungee Cord (or equivalent)
•
5/16” Socket Wrench
•
Side Cutters
•
9/16” Socket Wrench
•
Floor Jack
•
7/32” Hex Allen Socket Wrench
•
Sawhorses (or equivalent)
•
2 x 4 (or equivalent).
•
3/8” Socket Wrench (with 6” extension)
Procedure:
Note
If you need more space to work under the base frame, place a 2 x 4 across the base frame rails and use a floor jack
to raise the base frame off the floor.
1. Unplug the bed power cord from the wall socket.
2. Using a 5/16” socket wrench, remove the five bolts holding the lower lift cover to the base and remove the cover.
3. Using a #2 Phillips screwdriver, remove the three screws holding the upper lift cover to the base. If you want, hold
the covers out of the way by using bungee cords (or the equivalent) to secure them to the litter top.
4. Remove the lift motor and capacitor (refer to Lift Motor and Capacitor Removal and Replacement procedure).
5. Remove lift potentiometer (refer to Lift Potentiometer Replacement and Adjustment procedure).
6. Using a 5/16” socket wrench, remove the cable clamps holding the power and sensor coil cords on top of the lift
housing assembly. Cut the cable ties and disconnect the coil cords from under the lift housing. The power and
sensor coil cords are now free of the lift housing assembly. Drape them up out of the way.
7. Using a 7/32” hex Allen socket, remove the two screws holding the lift screws to the header crossbar plate.
8. Lift the litter top up and support it about 6” above the lift screws with sawhorses or the equivalent.
9. Under the base, using a 9/16” socket, remove the four nuts (A) holding the lift housing to the base.
10. Lift up and out on the lift housing assembly to remove it from the base.
CAUTION
To ensure proper reattachment of the power and sensor coil cords, refer to the Power and Sensor Coil Cord Replacement
procedure.
Refer to the Lift Potentiometer Replacement And Adjustment procedure for reattachment of the lift potentiometer.
11. Reverse the above steps to reinstall the lift housing assembly after service is completed.
Note
The procedure for lift housing removal and replacement is the same for both ends of the bed.
Service Information
Summary of Contents for EPIC II
Page 2: ......
Page 71: ...Return To Table of Contents www stryker com 2030 809 002 REV D 71 Notes ...
Page 77: ...Return To Table of Contents www stryker com 2030 809 002 REV D 77 Base Assembly and Options ...
Page 78: ...Return To Table of Contents 78 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 80: ...Return To Table of Contents 80 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 96: ...Return To Table of Contents 96 2030 809 002 REV D www stryker com Notes ...
Page 106: ...Return To Table of Contents 106 2030 809 002 REV D www stryker com Litter Assembly ...
Page 107: ...Return To Table of Contents www stryker com 2030 809 002 REV D 107 Litter Assembly ...
Page 108: ...Return To Table of Contents 108 2030 809 002 REV D www stryker com Litter Assembly ...
Page 109: ...Return To Table of Contents www stryker com 2030 809 002 REV D 109 Litter Assembly ...
Page 110: ...Return To Table of Contents 110 2030 809 002 REV D www stryker com Litter Assembly ...
Page 111: ...Return To Table of Contents www stryker com 2030 809 002 REV D 111 Litter Assembly ...
Page 112: ...Return To Table of Contents 112 2030 809 002 REV D www stryker com Litter Assembly ...
Page 113: ...Return To Table of Contents www stryker com 2030 809 002 REV D 113 Litter Assembly ...
Page 120: ...Return To Table of Contents 120 2030 809 002 REV D www stryker com Notes ...
Page 171: ...Return To Table of Contents www stryker com 2030 809 002 REV D 171 Foot End Siderail Assembly ...
Page 226: ......
Page 227: ......