Problem / Failure
Recommended Action
No bed down motion.
A. Enter diagnostics, (Refer to Scale System Diagnostics
and Calibration section) and press bed down.
a. If motion is not present, verify there is a two pin
shunt present on connector Y, closest to the center
of the bed, if not, install shunt.
1. Test bed down motion, if motion is present
then go to step D.
b. If motion is present, reburn lift potentiometer
limits, Refer to Lift Potentiometer Replacement
and Adjustment section for procedure.
B. Check for 5 VDC on TP 9 (HL) and TP 7 (FL) referencing
ground test point while pressing bed down.
a. If 5VDC is present, go to step C.
b. If 5VDC is not present, replace CPU board.
C. Check for 120VAC power on connector N (HL) and G
(FL), pin 1 (white) and pin 3 (black), of the CPU board,
while pressing bed motion up.
a. If voltage is not present, replace CPU board.
b. If voltage is present:
1. Verify the motors are running, if so, replace
lift couplers.
2. If motors are not running, check voltage at
motor connection.
3. If voltage is present at motor, check capacitors
or motors.
D. Verify bed function and return to service.
No bed up motion.
A. Enter diagnostics, (Refer to Scale System
Diagnostics and Calibration section) and press
bed up.
a. If motion is not present, go to step B.
b. If motion is present, reburn lift potentiometer
limits, refer to Lift Potentiometer Replacement
and Adjustment section for procedure.
B. Check for 5VDC on TP 10 (HL) and TP 8 (FL) on
the CPU board referencing ground test point while
pressing bed up.
a. If 5VDC is present, go to step C.
b. If 5VDC is not present, replace CPU board.
C. Check for 120VAC power on connector N (HL) and
G (FL), pin 1 (white) and pin 6 (red), of the CPU
board while pressing bed motion up.
a. If voltage is not present, replace CPU board.
b. If voltage is present:
1. Verify the motors are running, if so, replace
lift couplers.
2. If motors are not running, check voltage at
motor connection.
3. If voltage is present at motor, check
capacitors or motors.
D. Verify bed function and return to service.
Troubleshooting Guide
Summary of Contents for EPIC II
Page 2: ......
Page 71: ...Return To Table of Contents www stryker com 2030 809 002 REV D 71 Notes ...
Page 77: ...Return To Table of Contents www stryker com 2030 809 002 REV D 77 Base Assembly and Options ...
Page 78: ...Return To Table of Contents 78 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 80: ...Return To Table of Contents 80 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 96: ...Return To Table of Contents 96 2030 809 002 REV D www stryker com Notes ...
Page 106: ...Return To Table of Contents 106 2030 809 002 REV D www stryker com Litter Assembly ...
Page 107: ...Return To Table of Contents www stryker com 2030 809 002 REV D 107 Litter Assembly ...
Page 108: ...Return To Table of Contents 108 2030 809 002 REV D www stryker com Litter Assembly ...
Page 109: ...Return To Table of Contents www stryker com 2030 809 002 REV D 109 Litter Assembly ...
Page 110: ...Return To Table of Contents 110 2030 809 002 REV D www stryker com Litter Assembly ...
Page 111: ...Return To Table of Contents www stryker com 2030 809 002 REV D 111 Litter Assembly ...
Page 112: ...Return To Table of Contents 112 2030 809 002 REV D www stryker com Litter Assembly ...
Page 113: ...Return To Table of Contents www stryker com 2030 809 002 REV D 113 Litter Assembly ...
Page 120: ...Return To Table of Contents 120 2030 809 002 REV D www stryker com Notes ...
Page 171: ...Return To Table of Contents www stryker com 2030 809 002 REV D 171 Foot End Siderail Assembly ...
Page 226: ......
Page 227: ......