i
BED Awareness System Error Codes
ERROR HANDLING
•
The error handling system provides a mechanism for capturing errors known to the product and providing feedback
to the operator through logged errors or message errors.
•
All Logged errors generate a generic blinking error message which will be displayed on the status screen in the
lower area: “Error XXX Call Service”, where XXX is the error code.
•
Errors will only be logged if the error log does not already contain the error.
•
In the event multiple errors are displayed on the status screen, the error code is used to identify the priority. The
lowest code has the highest priority.
•
All logged errors and message errors are persistent until the power is cycled, or the condition is fixed.
•
The details for each logged error and error message are detailed in the table below.
LOGGED ERRORS
Error
Code
Name
Description
Type
Affected Systems
100
Not Used
Highest Priority Errors
102
StuckSafetyRelay
This error is checked while the product is
“On”. If the safety relay signal is “On” and
there is no button press on the product then
set this error after 1 minute.” Note: This should
not be checked when in Auto-Cycle Mode.
Logged
Motion
103
InputTimer
If any footboard motion or side rail/pendant
input is sensed to be “On” continuously for a
total of 5 minutes then provide an exception
message. Note: This should not be checked
when in Auto-Cycle Mode.
Logged
Motion
200
Not Used
Scale System
201
HL-Load Cell
Patient Head Left Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
202
HR-Load Cell
Patient Head Right Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
203
FL-Load Cell
Patient Foot Left Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
204
FR-Load Cell
Patient Foot Right Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
206
CS5533Fatal
Non-Recoverable Communication Error.
CS5533 Fatal occurs when the CPU fails
to communicate to the External ADC for the
Load cells and cannot recover.
Logged
Scale, Bed Exit,
i
BED Awareness
207
ZeroFail
Zero Fail error occurs when the system can-
not provide a stable zero value for three con-
secutive attempts at zeroing the product.
Logged
Scale, Bed Exit,
i
BED Awareness
Summary of Contents for EPIC II
Page 2: ......
Page 71: ...Return To Table of Contents www stryker com 2030 809 002 REV D 71 Notes ...
Page 77: ...Return To Table of Contents www stryker com 2030 809 002 REV D 77 Base Assembly and Options ...
Page 78: ...Return To Table of Contents 78 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 80: ...Return To Table of Contents 80 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 96: ...Return To Table of Contents 96 2030 809 002 REV D www stryker com Notes ...
Page 106: ...Return To Table of Contents 106 2030 809 002 REV D www stryker com Litter Assembly ...
Page 107: ...Return To Table of Contents www stryker com 2030 809 002 REV D 107 Litter Assembly ...
Page 108: ...Return To Table of Contents 108 2030 809 002 REV D www stryker com Litter Assembly ...
Page 109: ...Return To Table of Contents www stryker com 2030 809 002 REV D 109 Litter Assembly ...
Page 110: ...Return To Table of Contents 110 2030 809 002 REV D www stryker com Litter Assembly ...
Page 111: ...Return To Table of Contents www stryker com 2030 809 002 REV D 111 Litter Assembly ...
Page 112: ...Return To Table of Contents 112 2030 809 002 REV D www stryker com Litter Assembly ...
Page 113: ...Return To Table of Contents www stryker com 2030 809 002 REV D 113 Litter Assembly ...
Page 120: ...Return To Table of Contents 120 2030 809 002 REV D www stryker com Notes ...
Page 171: ...Return To Table of Contents www stryker com 2030 809 002 REV D 171 Foot End Siderail Assembly ...
Page 226: ......
Page 227: ......