FOWLER POTENTIOMETER REPLACEMENT (CONTINUED)
Service Information
9. Unplug the bed power cord from the wall socket.
10. Press and hold the “Knee Lockout” and “Bed Motion Lock” buttons on the footboard.
11. Plug the power cord into the wall socket. Release the two buttons and the “Knee Lockout” LED should be flashing.
This indicates the bed is in the correct calibration mode.
12. Using the footboard controls, run the Fowler up to 90°. Press and hold the button on the footboard to lock out the
siderail controls for the back until the padlock LED flashes. Release the button.
13. Using the footboard controls, run the Fowler down to 0°. Press and hold the button on the footboard to lock out the
siderail controls for the knee until the padlock LED flashes. Release the button and unplug the bed power cord to
complete the “burn-in” mode.
14. Plug the bed power cord into the wall socket and verify the back limits are set properly before returning the bed
to service.
FOWLER POTENTIOMETER “BURN-IN” PROCEDURE
Note
It requires
two people
to enable the diagnostics mode for the bed.
1. Unplug the bed power cord from the wall socket.
2. On the footboard control panel, hold down the bed motion lock button and the button to lock out the siderail
controls for the knee. While still holding the buttons, plug the bed power cord into the wall socket. Release the
footboard buttons. The siderail control lights LED should be flashing to indicate the bed is in diagnostics mode.
3. Using the footboard controls, run the Fowler up to 90°. Press and hold the button on the footboard to lock out the
siderail controls for the back until the padlock LED flashes. Release the button.
4. Using the footboard controls, run the Fowler down to 0°. Press and hold the button on the footboard to lock out the
siderail controls for the knee until the padlock LED flashes. Release the button..
FOWLER POTENTIOMETER “BURN-IN” PROCEDURE (
I
BED AWARENESS OPTION)
Note
It requires
two people
to enable the diagnostics mode for the bed.
1. Unplug the bed power cord from the wall socket.
2. On the footboard control panel, hold down the Patient Fowler Lock button and the Patient Bed Up/Down Lock
button. While still holding the buttons, plug the bed power cord into the wall socket. Release the footboard
buttons.
3. The message “Set Fowler to 1
0
” will be displayed on the screen.
4. Using the footboard or siderail controls, set the Fowler to 1
0
.
5. Press and Hold down the Enter button. The message “Release Button” will be displayed on the screen.
6. After releasing button, the message “Set Fowler to 30
0
” will be displayed on the screen.
7. Using the footboard or siderail controls, set the Fowler to 30
0
.
8. Press and Hold down the Enter button. The message “Release Button” will be displayed on the screen.
9. After releasing button, the message “Set Fowler to 45
0
” will be displayed on the screen.
10. Using the footboard or siderail controls, set the Fowler to 45
0
.
11. Press and Hold down the Enter button. The message “Release Button” will be displayed on the screen.
12. After releasing button, the message “Set Fowler to 90
0
” will be displayed on the screen.
13. Using the footboard or siderail controls, set the Fowler to 90
0
.
14. Press and Hold down the Enter button. The message “Release Button” will be displayed on the screen.
15. After releasing button, the message “Fowler Calibration Complete” will be displayed on the screen.
Summary of Contents for EPIC II
Page 2: ......
Page 71: ...Return To Table of Contents www stryker com 2030 809 002 REV D 71 Notes ...
Page 77: ...Return To Table of Contents www stryker com 2030 809 002 REV D 77 Base Assembly and Options ...
Page 78: ...Return To Table of Contents 78 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 80: ...Return To Table of Contents 80 2030 809 002 REV D www stryker com Base Assembly and Options ...
Page 96: ...Return To Table of Contents 96 2030 809 002 REV D www stryker com Notes ...
Page 106: ...Return To Table of Contents 106 2030 809 002 REV D www stryker com Litter Assembly ...
Page 107: ...Return To Table of Contents www stryker com 2030 809 002 REV D 107 Litter Assembly ...
Page 108: ...Return To Table of Contents 108 2030 809 002 REV D www stryker com Litter Assembly ...
Page 109: ...Return To Table of Contents www stryker com 2030 809 002 REV D 109 Litter Assembly ...
Page 110: ...Return To Table of Contents 110 2030 809 002 REV D www stryker com Litter Assembly ...
Page 111: ...Return To Table of Contents www stryker com 2030 809 002 REV D 111 Litter Assembly ...
Page 112: ...Return To Table of Contents 112 2030 809 002 REV D www stryker com Litter Assembly ...
Page 113: ...Return To Table of Contents www stryker com 2030 809 002 REV D 113 Litter Assembly ...
Page 120: ...Return To Table of Contents 120 2030 809 002 REV D www stryker com Notes ...
Page 171: ...Return To Table of Contents www stryker com 2030 809 002 REV D 171 Foot End Siderail Assembly ...
Page 226: ......
Page 227: ......