![Stryker BERTEC GOBED PLUS FL20E Operation Manual Download Page 8](http://html.mh-extra.com/html/stryker/bertec-gobed-plus-fl20e/bertec-gobed-plus-fl20e_operation-manual_1385870008.webp)
GO
BED
+
Operations Manual
8
1.7 SET-UP PROCEDURE
CHECKLIST
It is important to ensure that the bed is working properly before it is put into service. The
following list will help ensure that each part of the bed is checked.
_____ Install the foot and head boards on the bed. Insert the foot board carefully so that the
board and the casing connectors fit in smoothly.
N
N
O
O
T
T
E
E
Boards can be fixed permanently by using two 1/4" flat washers/1/4" spring washers/1/4-20 x 1
3/4" hexagon screws for the head board and two 1/4" flat washers/1/4" spring washers/1/4-20 x
2 1/4" hexagon screws for the foot board. Simply insert a screw and the two different washers in
factory installed nuts (refer to drawing 20-0130L, attached to the parts list L20-001, for the
illustration of the nut (part VE20A1N -ref. no. 17)) accessible through holes located on the inner
sides of the foot end and head end casings.
_____ Plug the power cord to the bed connector at the head end of the bed and into a properly
grounded hospital grade wall outlet. Turn on the power switch loacted on the power
connecter.
The bed is equipped with a hospital grade plug for protection against shock hazard. It must be
plugged directly into a properly grounded receptacle. Grounding reliability can be achieved only
when a hospital grade receptacle is used.
_____ Depress the red “TOTAL BRAKE” side of the pedal at both sides of the bed fully down to
set the brakes (see page 12). Push on the bed to ensure it is immobilized. Toggle the
pedal to the neutral position and ensure the brakes are released.
_____ Depress the green “AXIAL STEER” side of the pedal at both sides of the bed fully down
to put the bed in the steer mode (see page 12). Ensure the 5th steer wheel is properly
engaged. Toggle the pedal to the neutral position and ensure the 5th wheel disengages.
_____ Ensure the siderails raise, lock in the up position and lower smoothly (see page 15).
_____ Run through each control of the foot board control panel (see page 17).
_____ Verify the optional Weigh and Bed Exit systems for proper operation (see page 18,
"Operating (Zeroing) the Scale…" and page 20 or 21, "To
activate
the Bed Exit…").
_____ Run through each function on both head end siderails (see page 16).
_____ Raise the bed to full up position and activate the Trendelenburg function (see page 17).
Ensure the head end lowers to the full down position.
_____ Raise the bed to the full up position and activate the reverse Trendelenburg function (see
page 17). Ensure the foot end lowers to the full down position.
_____ Verify the optional CPR hold-to-run release mechanism. Raise the Fowler to full up.
Using the CPR release handle (see page 14), lower the Fowler gradually to flat position
by pulling, holding and releasing the handle several times. Ensure the Knee Gatch (if
raised) also start flattening when the Fowler is completely down. Verify that the Fowler
actuator resets itself when the Fowler is completely lowered. Note that during the Fowler
motor reset period, the bed functions are not available.
_____ Verify the following optional equipments for proper operation: 120V auxiliary outlet, night
light, two or three function pendant control, emergency crank operation, etc.
If any problems are found during bed set-up, contact our Technical Service department (see
section 1.2).
W
W
A
A
R
R
N
N
I
I
N
N
G
G
Summary of Contents for BERTEC GOBED PLUS FL20E